31
EN
Operating conditions:
10 - 40 °C / 50 - 104 °F
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 90% relative maximum humidity
Weight:
130 g (including batteries)
Dimensions:
80 x 70 x 70 mm
Measuring procedure:
oscillometric, corresponding to Korotkoff method:
phase I systolic, phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
2 x 1.5 V Alkaline-Batteries; size AAA
Battery life:
approx. 320 measurements (new batteries)
IP class:
IP 20
Reference to standards:
EN 1060-1 /-3 /-4; IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10,000 measurements; accessories: 2 years
11. Technical Specifications
Summary of Contents for aponorm Mobil Soft Control
Page 1: ...Gebrauchsanweisung Handgelenk MOBIL SOFT CONTROL DE EN FR TR RU NEU...
Page 2: ...aponorm Mobil Soft Control Kurzanleitung...
Page 3: ...aponorm Mobil Soft Control 1 5 1 6 4 3 2 7 8 9 10 12 13 14 11 15...
Page 5: ...1 DE...
Page 55: ...51 RU 1...
Page 57: ...53 RU A B 3 1 5 6 2 3 4 5 3 2...
Page 58: ...54 1 2 5 3 4 5 6 13 5 21 5 7 8 9 10 14 11 7 8 9 12 1 3...
Page 59: ...55 RU 5 11 6 1 4 13...
Page 60: ...56 5 M17 17 M 30 CL CL 6...
Page 61: ...57 RU 15 15 1 3 2 3 2 7...
Page 62: ...58 2 1 5 AAA NiMH...
Page 63: ...59 RU ERR 3 1 8 ERR 1 ERR 2 ERR 3 ERR 5 HI 300 200 5 LO 40...
Page 64: ...60 1 9...
Page 65: ...61 RU 2 aponorm by microlife...
Page 66: ...62 5 aponorm a by microlife 10...
Page 68: ...64 93 42...