17

EN

Using rechargeable batteries

You can also operate this device using rechargeable batteries.

  Only use «NiMH» type reusable batteries.

  Batteries must be removed and recharged when the flat

battery symbol appears. They should not remain inside the

device as they may become damaged (total discharge as a

result of low use of the device, even when switched off).

  Always remove the rechargeable batteries if you do not

 

  intend to use the device for a week or more.

  Batteries cannot be charged in the blood pressure monitor.

Recharge batteries in an external charger and observe the

information regarding charging, care and durability.

7.  Using a Mains adapter

You can operate this device using the 

aponorm

®

 by microlife 

mains adapter (DC 6V, 600mA).

  Only use the 

aponorm

®

 by microlife mains adapter 

 

  available as an original accessory appropriate for your    

 

  supply voltage.

  Ensure that neither the mains adapter or the cable are

   damaged.
1.  Plug the adapter cable into the mains adapter socket 

4

 in   

 

the blood pressure monitor.

2.  Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is 

consumed.

8.  Error Messages

If an error occurs during the measurement, the measurement is

interrupted and an error message, e.g. «

ERR 3

», is displayed.

* Please consult your doctor, if this or any other problem occurs

 repeatedly.

  If you think the results are unusual, please read through   

 

  the information in «Section 1.» carefully.

Error

«ERR 1»

«ERR 2»

«ERR 3»

«ERR 5» 

«Hi»

«LO»

Description

Signal too

weak

Error signal

No pressure

in the cuff

Abnormal

result

Pulse or cuff

pressure too

high

Pulse too 

low

Potential cause and remedy 

The pulse signals on the cuff are too

weak. Re-position the cuff and repeat 

the measurement.*

During the measurement, error signals

were detected by the cuff, caused for

instance by movement or muscle 

tension.Repeat the measurement, 

keeping your arm still.

An adequate pressure cannot be gene-

rated in the cuff. A leak may have

occurred. Check that the cuff is cor-

rectly connected and is not too loose. 

Replace the batteries if necessary. 

Repeat the measurement.

The measuring signals are inaccurate 

and no result can therefore be dis-

played. Read through the checklist for 

performing reliable measurements and 

then repeat the measurement.*

The pressure in the cuff is too high 

(over 300 mmHg) OR the pulse is too 

high (over 200 beats per minute). 

Relax for 5 minutes and repeat the 

measurement.*

The pulse is too low (less than 40 

beats per minute). Repeat the measu-

rement.*

Summary of Contents for Aponorm Basis Control

Page 1: ...Gebrauchsanweisung B a s i s Oberarm DE 2 RU 26 EN 10 TR 18 ...

Page 2: ...n automa tisch gespeichert Schließen Sie die Manschette an Starten Sie die Messung mit einem Druck auf Start Stopp Sie können das Gerät wieder verriegeln in dem Sie den Schalter nach oben schieben Entriegeln Sie das Gerät indem Sie den Schalter an der Seite nach unten schieben Während der Messung weder sprechen bewegen essen trinken oder rauchen Vor jeder Messung sollten Sie einige Minuten entspan...

Page 3: ...aponorm Basis aponorm by 3 DE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 ...

Page 4: ...oration with physicians and clinical tests carried out proving its measurement accuracy to be of a very high standard Please read through these instructions carefully so that you un derstand all functions and safety information We want you to be happy with your aponorm by microlife product If you have any questions problems or want to order spare parts please contact aponorm by microlife Customer ...

Page 5: ...manently high blood pressure values can damage your health and must be treated by your doctor Always discuss your values with your doctor and tell him her if you have noticed anything unusual or feel unsure Never rely on single blood pressure readings There are several causes of excessively high blood pressure values Your doctor will explain them in more detail and offer treatment where appropriat...

Page 6: ... year by pressing either the 11 or the 10 button To confirm and then set the month press the time button 8 2 Press the 11 or the 10 button to set the month Press the time button 8 to confirm and then set the day 3 Follow the instructions above to set the day hour and minutes 4 Once you have set the minutes and pressed the time button the date and time are set and the time is displayed 5 If you wan...

Page 7: ...automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 8 When the correct pressure is reached the pumping stops and the pressure falls gradually If the required pressure was not reached the device will automatically pump some more air into the cuff 9 During the measurement the pulse indicator 20 flashes in the di...

Page 8: ...press START STOP button 1 while CL is flashing Individual values cannot be cleared 6 Battery Indicator and Battery change Low battery When the batteries are approximately empty the battery symbol 17 will flash as soon as the device is switched on partly filled battery displayed Although the device will continue to measure reliably you should obtain replacement batteries Flat battery replacement Wh...

Page 9: ...Messages If an error occurs during the measurement the measurement is interrupted and an error message e g ERR 3 is displayed Please consult your doctor if this or any other problem occurs repeatedly If you think the results are unusual please read through the information in Section 1 carefully Error ERR 1 ERR 2 ERR 3 ERR 5 HI LO Description Signal too weak Error signal No pressure in the cuff Abn...

Page 10: ...ff is included for the functional guarantee only tightness for 2 years Please contact aponorm by microlife Service see foreword aponorm by 18 9 Safety Care Accuracy Test and Disposal Safety and protection This device may only be used for the purposes described in this instruction The manufacturer cannot be held liable for damage caused by incorrect application This device comprises sensitive compo...

Page 11: ...thod Phase I systolic Phase V diastolic Measurement range 20 280 mmHg blood pressure 40 200 beats per minute pulse Cuff pressure display 0 299 mmHg range Resolution 1 mmHg Static accuracy pressure within 3 mmHg Pulse accuracy 5 of the readout value Voltage source 4 x 1 5 V Batteries size AAA Mains adapter DC 6 V 600 mA optional Reference to EN 1060 1 3 4 IEC 60601 1 standards IEC 60601 1 2 EMC Thi...

Page 12: ...aponorm by 37 37 ...

Page 13: ...rien Nr Seriennummer Серийный номер Numer seryjny Sorozatszám Număr de serie Výrobné číslo Seri Numarası Date of Purchase Date d achat Fecha de compra Data da compra Kaufdatum Datum van aankoop Дата покупки Data zakupu Vásárlás dátuma Data cumpărării Datum nákupu Dátum kúpy Satın Alma Tarihi Specialist Dealer Revendeur Vendedor especializado Revendedor autorizado Apotheker Specialist Dealer Специа...

Page 14: ...ife Garantie Ihr neues Blutdruckmessgerät ist ein medizinisches Präzisionsgerät der Spitzenklasse Dafür verbürgen wir uns mit unserem guten Namen und unserer 5 Jahres Garantie Die genauen Garantie Bedingungen finden Sie auf Seite 9 der Gebrauchsanweisung ...

Page 15: ...darf GmbH Co KG Am Fichtenstrauch 6 10 56204 Hillscheid www aponorm de www wepa dieapothekenmarke de Hersteller Microlife AG Espenstraße 139 CH 9443 Widnau www microlife ch Hilfspositionsnummer 21 28 01 2104 Bestell Nr 46928 02391246 ...

Reviews: