background image

7

EN

A7 Touch BT

a result of low use of the instrument, even when switched off).

  Always remove the rechargeable batteries if you do not intend 

to use the instrument for a week or more.

  Batteries cannot be charged in the blood pressure monitor. 

Recharge batteries in an external charger and observe the 

information regarding charging, care and durability.

11. Using a Mains Adapter

You can operate this instrument using the Microlife mains adapter

(DC 6V, 600mA).

  Only use the Microlife mains adapter available as an original 

accessory appropriate for your supply voltage.

  Ensure that neither the mains adapter nor the cable are 

damaged.

1.   Plug the adapter cable into the mains adapter socket 

4

 in the 

blood pressure monitor.

2.   Plug the adapter plug into the wall socket.

When the mains adapter is connected, no battery current is

consumed.

12. Error Messages

If an error occurs during the measurement, the measurement is

interrupted and an error message, e.g. «ERR 3», is displayed.

* Please consult your doctor, if this or any other problem occurs

Error

Description Potential cause and remedy

«ERR 1» Signal too

weak

The pulse signals on the cuff are too 

weak. Re-position the cuff and repeat 

the measurement.*

«ERR 2»

BS

Error signal During the measurement, error signals 

were detected by the cuff, caused 

for instance by movement or muscle 

tension.

Repeat the measurement, keeping your 

arm still.

«ERR 3»

CT

No pressure

in the cuff

An adequate pressure cannot be 

generated in the cuff. A leak may have 

occurred. Check that the cuff is correctly 

connected and is not too loose. Replace 

the batteries if necessary. Repeat the 

measurement.

«ERR 5» Abnormal

result

The measuring signals are inaccurate

and no result can therefore be displayed.

Read through the checklist «Section 4.»for 

performing reliable measurements and 

then repeat the measurement.*

«ERR 6» AFIBsens

TM

/ MAM Mode

There were too many errors during the 

measurement in AFIBsens

TM

 / MAM 

mode, making it impossible to obtain a 

final result. Read through the checklist 

«Section 4.»

 for performing reliable 

measurements and then repeat the 

measurement.*

«HI»

Pulse or cuff

pressure too

high

The pressure in the cuff is too high (over 

300 mmHg) OR the pulse is too high 

(over 200 beats per minute). Relax for 5 

minutes and repeat the measurement.*

«LO»

Pulse too

low

The pulse is too low (less than 40 beats 

per minute). Repeat the measurement.*

Summary of Contents for A7 Touch BT

Page 1: ...EN 1 Microlife BP A7 Touch BT A7Touch BT...

Page 2: ...Microlife BP A7 Touch BT Start Stop _...

Page 3: ...r want to order spare parts please contact Microlife Customer Service Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country Alternatively visit the Inter...

Page 4: ...Cleaning the cuff Accuracy test Disposal 14 Guarantee 15 Technical Specifications 1 ImportantFactsaboutBloodPressureandSelf Measurement Blood pressure is the pressure of the blood flowing in the arter...

Page 5: ...t and pump blood Atrial fibrillation occurs when rapid disorganized electrical signals are present in the heart s two upper chambers called the atria causing them to contract quickly fast and irregula...

Page 6: ...on The first person to measure should select user 1 Select the measuring mode standard or AFIBsensTM mode This device enables you to select either standard standard single measurement or AFIBsensTM MA...

Page 7: ...ssure sensation 5 AppearanceoftheAtrialFibrillationIndicatorforearly Detection onlyinAFIB MAMmode This device is able to detect atrial fibrillation AFIB This symbol BO indicates that atrial fibrillati...

Page 8: ...values 1 Firstly unlock the device then select either 1 or 2 with the user pad 9 2 Hold down the M pad AK until CL appears and then release the button 3 Touch the M pad while CL is flashing to permane...

Page 9: ...ERR 1 Signal too weak The pulse signals on the cuff are too weak Re position the cuff and repeat the measurement ERR 2 BS Error signal During the measurement error signals were detected by the cuff c...

Page 10: ...ections of this booklet Ensure that children do not use the instrument unsupervised some parts are small enough to be swallowed Device care Clean the instrument only with a soft dry cloth Cleaning the...

Page 11: ...urce 4 x 1 5 V Batteries size AAA Mains adapter DC 6V 600 mA optional Battery lifetime approx 400 measurements using new batteries IP Class IP20 Reference tostandards EN 1060 1 3 4 IEC 60601 1 IEC 606...

Page 12: ...w microlife com Asia Microlife Corporation 9F 431 RuiGuang Road NeiHu Taipei 11492 Taiwan R O C Tel 886 2 87971288 Fax 886 2 87971283 Email service microlife com tw www microlife com North Central Sou...