![Microcyber CS-TT106H-V2.0 User Manual Download Page 18](http://html1.mh-extra.com/html/microcyber/cs-tt106h-v2-0/cs-tt106h-v2-0_user-manual_1786196018.webp)
Temperature Transmitter | NCS TT106 | Page
Email | [email protected]
http://www.microcyber-fieldbus.com
13
then the value of the cold-end temperature compensation
is measured by the external Pt100.
Channel status: Display sensor channel status (open circuit,
short circuit & etc.).
Sensor channel value: Display the original value of sensor
channel.
Cold end temperature: Display the current cold end
temperature.
R0 Correction Factor: Correction of the error of the sensor
itself (range 0.9-1.1).
Two-wire zero calibration: When the temperature
transmitter is connected to RTD in 2-wire mode,to avoid
the error caused by the resistance on the cable, the sensor
end can be short connected, and then the zero calibration
button can be executed to eliminate the error caused by
the resistance on the cable.
TC Calibration: Calibration of various thermocouple ranges of
transmitters is carried out in the factory (only applicable to
manufacturer-level users).
RTD calibration: All kinds of resistance ranges of transmitter
are calibrated out of the factory (only for
manufacturer-level users).
。
Restore factory default value: Click this button, all data will
be restored to the default factory status.
Save as factory value: Click this button to save the current
configuration as factory value. When you click the “Restore