8
1. Rinse excess soil from top
2. Using brush, apply detergent solution to all
surfaces
3. Rinse under clear running water
Desinfection
Not recommended
Drying
Do not exceed 134º C
Maintenance, inspection and testing
Discharge damaged products
Packaging
Only pack individually
Sterilization
Vacuum autoclave, minimum 18 minutes at 134º C,
pressure 2.2 bar. Do not exceed 134º C.
Storage
No particular requirements
4. Labelling
Manufacturer
See instruction manual
Serial number
Medical Electrical Equipment Directive,
DS/EN 60601-1:2006
Electric shock protection, Type B
Medical Devices Directive, MDD Directive
93/42/EEC
Medical device, Class I
Canadian Standards Association, CSA
Approved for use in Canada and the USA
Waste Electrical and Electronic Equipment
Directive,
WEEE Directive 2002/96/EC
Please return the product for recycling
Accessories
HEROfill
®
Obturator References
HEROfill
®
Obturator n°20 - Ref. 20138501 N
HEROfill
®
Obturator n°25 - Ref. 20138502 N
HEROfill
®
Obturator n°30 - Ref. 20138503 N
HEROfill
®
Obturator n°35 - Ref. 20138504 N
HEROfill
®
Obturator n°40 - Ref. 20138505 N
HEROfill
®
Obturator n°45 - Ref. 20138506 N
Accessories References
HEROfill
®
Verifier n°20 - Ref. 20139501 N
HEROfill
®
Verifier n°25 - Ref. 20139502 N
HEROfill
®
Verifier n°30 - Ref. 20139503 N
HEROfill
®
Verifier n°35 - Ref. 20139504 N
HEROfill
®
Verifier n°40 - Ref. 20139505 N
HEROfill
®
Verifier n°45 - Ref. 20139506 N