Page 21
Technical Data
thermoIMAGER TIM
2.7
Microscope Lenses
Model
TIM 450
TIM 640
Optical resolution
382 x 288 Pixel @ 80 Hz
640 x 480 Pixel @ 32 Hz
640 x 120 Pixel @ 125 Hz
Temperature ranges (scalable)
-20 to 100°C, 0 to 250 °C, (20)150 to 900 °C
1)
Spectral range
7.5 to 13 μm
Frame rate
80 Hz (switchable to 27 Hz)
125 Hz (switchable to 32 Hz)
System accuracy
± 2 °C or ± 2 %, whichever is greater
Microscope lens (FOV)
10° x 8° (F = 1.1) / f = 44 mm
12° x 9° (F = 1.1) / f = 44 mm
Smallest spot size (IFOV)
42 μm
28 μm
MFOV
85 μm
2)
Focus adjustment
80 to 100 mm
Thermal sensitivity (NETD)
90 mK
120 mK
Detector
FPA – uncooled micro bolometer
Outputs/digital
USB 2.0
Standard process interface (PIF)
0 - 10 V input, digital input (max. 24 V), 0 - 10 V output
Industry process interface (IPIF)
2 x 0 -10 V input, digital input (max. 24 V), 3 x 0 -10 V output,
3 x Relais (0 - 30 V / 400 mA), fail-safe relay
Cable length (USB)
1 m (standard), 3 m, 5 m, 10 m, 20 m
Power supply
USB powered
Tripod mount
¼-20 UNC
Protection class
IP 67
Ambient temperature range
0 °C to 70 °C
0 °C to 50 °C
Storage temperature
-40 °C to 70 °C
Relative humidity
20 to 80 %, non-condensing