scanCONTROL 25xx

Trigger, Encoder, Mode Switching

The switching inputs of the multifunction port can either be used for encoder input, 
trigger input or to load previously stored user modes.

The signal levels are switchable for all switching inputs between LLL (TTL logic) and 
HLL (HTL logic):

 -

LLL level: Low 0 V … 0.8 V, high 2.4 V … 5 V, internal pull-up 10 kΩ connected to 5 V

 - HLL-Pegel: Low 0 V … 3 V, high 11 V … 24 V (up to 30 V are permissible), internal 

pull-up  

10 kΩ connected to 24 V

 -

Pulse duration: ≥ 5 µs

External Laser Switch-off, Optional

The external laser switch-off is implemented as a hardware solution and is a top priority. 
The laser can also be switched off by software. 

This function is offered by sensors with the /SI option.

 

Use a serial key switch in the control circuit to switch off the laser

With standard sensors, connecting the supply voltage activates the laser light source in 
the sensor. 
Wiring details are available in the operating instructions, Chap. 5.2.6.

Ethernet Connection, Standard Connection to PC

RJ45 connector

8-pin screw connector (sensor side)

Pin no. Color of stranded hook-

up wire SCR3000A-x

Pin no. 10BaseT,  

100BaseTX

1000BaseT

1

White (orange)

5

Tx+

D1+

2

Orange

6

Tx-

D1-

3

White (green)

8

Rx+

D2+

4

Blue

1

D3+

5

White (blue)

2

D3-

6

Green

7

Rx-

D2-

7

White (brown)

3

D4+

8

Brown

4

D4-

1

View on pin side 
of male cable 
connector

2

3

4

5

6

7

8

1

View on solder side  
(cable) of screw connec-
tor (A-coded)

 - The sensor supports an automatic, sensor-specific IP address in the link-local net 

(169.254.x.x). No collision detection is implemented. This is also the default setting.

 - The sensor supports DHCP. This setting is activated by default and has priority over 

search the link-local net. 

 - The scanCONTROL 25xx sensor supports Power over Ethernet.

Summary of Contents for scanCONTROL 25 Series

Page 1: ...xposure of sensor to aggressive media deter gents cooling emulsions Damage to or destruction of the sensor 2 Notes on CE Marking The following apply to the scanCONTROL 25xx measuring system EU Directi...

Page 2: ...th warning labels are covered over when the unit is installed the user must ensure that supplementary labels are attached Laser Class 2M The sensors fall within laser class 2M IIM The laser is operate...

Page 3: ...DESCRIBED IN LASER NOTICE NO 56 DATED MAY 8 2019 Only for USA Class 3B laser sensors are notifiable and a laser protection officer is required Mark the laser area recognizable and everlasting During o...

Page 4: ...urple Switching input In3 GND In3 7 Black Ground connection In3 Screen Housing Black Not electrically connected to GND 2 3 4 5 6 7 8 9 1 10 11 12 GND galvanically isolated from In1 In2 In3 RS422 Laser...

Page 5: ...cuit to switch off the laser With standard sensors connecting the supply voltage activates the laser light source in the sensor Wiring details are available in the operating instructions Chap 5 2 6 Et...

Page 6: ...Meaning Green Measurement is active Green flashes long Data transmission is active Green flashes short Controller is accessing laser Red flashes Error code 6 System Requirements scanCONTROL Configura...

Page 7: ...Close unused connection plugs with the included protective caps Turn on the power supply 24 VDC Please wait until the scanCONTROL measuring system is recognized by the PC This may take a few seconds N...

Page 8: ...Data Profile data of scanCONTROL can be accessed via GigEVision and GenICam for digital cameras via Ethernet interface SDK for fast integration into applications C C C and others Please refer to the r...

Page 9: ...or can be transmitted to a PLC via one of these fieldbus systems Furthermore all scanCONTROL sensor settings can be set via the 2D 3D Gateway e g Laser on off or User Mode Sensor parameterization Supp...

Page 10: ...2 5 3 0 5 7 LINK ACT MS I O ETHERNET NS 8 7 6 5 4 3 2 1 N O LINK ACT 1 2 X 1 X 2 8 1 ON 24V X3 0V 0 WBM 255 DHCP OU fieldbus coupler L1 L2 L3 N PE OU Filter module OU Fieldbus coupler 13 14 C D B A 75...

Page 11: ...al output terminals need a field supply of 5 VDC or 24 VDC depending on the module type After mounting the modules install the required wiring Connect the System supply out contacts of the OU Filter m...

Page 12: ...and free cable ends Calibration protocol assembly instructions 2 Protective caps 14 Additional Information Please refer to the enclosed online documentation the Status and Error Messages and Notes sec...

Reviews: