Page 68
Operation
optoCONTROL 1200/1201
Example with offset
Analog signal with free beam 10 V
With 100 % shading the offset is 0.02 V.
With free beam, set the analog signal to 10.02 V.
x
[mm] =
MR
[mm] * (10 V + offset -
V
OUT
)
10 V
x
[mm] =
10 mm * (10 V + 0.02 V - 5.02 V)
10 V
x
[mm] =
5 mm