Installation and Assembly
i
IMPORTANT!
The tuned cables may not
be shortened because this
changes the capacity and
the adjustment of the
measuring system!
The thicker cable end must
be connected with the signal
conditioning electronics.
4.3 Sensor Cables
Do not kink the cables - the minimum bending radius is 10 mm.
Lay the cable such that no sharp-edged or heavy objects can affect the
cable sheath.
Make the connection between the sensor and controller using the integral
sensor cable, the transition board and the sensor cable C3/1. Connect the
sensor cable to the controller. Check the plugged connections for firm sea
ting.
The integral sensor cable has a length of 0.5 m. The sensor cable extension
C3/1 has a length of 3 m.
The cables of one type are tuned to the same capacitance at the factory. This
may result in deviations in length of +15 % in relation to the nominal length.
Fig. 4.3: Measuring against the injection needle
Injection needle
Injection body
Mounting screw
Sensor LS04
0.5
(.020)
3
(.118)
ø4 (.16)
0.2
(.008)
M4
x
0.35
0.61
(.024)
2.71
(.107)