Page 21
Initial Operation
IMS5x00
Material Selection
The refractive index needs to be corrected in the controller for an exact distance or thickness measurement. Only
air may be present between the sensor face and the measuring object (
Material Infront
); other media such as
water or alcohol are not permissible.
Dicke
Controller IMC5400, IMC5600
Abstand
Layer 1
Material
Infront
Peak 1
Switch to material selection by going to
Settings
>
Data recording
.
Peak 1
Peak 2
Peak A
1
Layer 1
Material
Infront
Assign the material according to the
target used.
You can edit or add to the material table. For a new material, a phase index and group refractive index is required.
Switch to the
Settings
>
Data recording
>
Link to material table
menu.
1) During a distance measurement on transparent objects, thickness peaks are displayed in addition to the distance
peaks, but the controller does not evaluate them. For the sake of clarity, thickness peaks are marked with letters.