Page 9
LEDs Controller
IMS5x00
LEDs Controller
Power on
Green
Operating voltage available
Status
Off
No error
If EtherCAT is active, meaning of the LED is conform with the Ether-CAT guidelines.
Intensity
Red
Signal in saturation
Yellow
Signal too low
Green
Signal ok
Range
Red
No target object, or target object outside the measuring range
Yellow
Target close to mid of measuring range
Green
Target within measuring range
SLED
Red
SLED of
Yellow
SLED warms up
Green
SLED ready for operation
Yellow flashing
SLED current outside the optimal value range
1
Pilot
Red
Pilot laser off
Green
Pilot laser on
Green
Pilot laser is alternately turned on and off,
if no target object or outside the measuring range
1) When measuring outside the optimum current value of the SLED, the controller will measure, but the measurement
accuracy may not be as specified.
The LED’s Intensity and Range flashes with their current color during a synchronization error.