background image

Operating Instructions

FSC1/7, FSC1000

PReliminary

Summary of Contents for FSC1/7

Page 1: ...Operating Instructions FSC1 7 FSC1000 PReliminary...

Page 2: ...nigbacher Strasse 15 94496 Ortenburg Germany Phone 49 0 8542 168 0 Fax 49 0 8542 168 90 email info micro epsilon com www micro epsilon com Layer thickness measuring device for insulators on metal and...

Page 3: ...packing Included in Delivery 10 3 2 Storage 10 4 Initial Operation 11 4 1 Dimensional Drawings 11 4 2 Battery Pack 13 5 FSC1 7 and FSC1000 Operation 14 5 1 Switch on Phase Warm up 14 5 2 General Contr...

Page 4: ...ement Data 30 5 8 3 Managing Measurement Data 31 5 8 4 Saving Exporting Measurement Data 31 5 9 Ending the Measuring Program Switching off the Controller 33 6 Liability for Material Defects 33 7 Clean...

Page 5: ...a tip for users Measurement Indicates hardware or a software button menu 1 2 Warnings Avoid shocks and impacts to the sensor Damage to or destruction of the sensor Avoid exposure of sensor and control...

Page 6: ...als and Part 15 Class C Low Power Communication Device Transmitter with the US Federal Communications Commission FCC This device complies with part 15 of the FCC Rules Operation is subject to the foll...

Page 7: ...used to measure the thickness of paint and other electrically insulating layers on substrates such as carbon fiber reinforced plastics CFRP CFRP with metallic lightning protection and metals The syst...

Page 8: ...f dielectric layers on CFRP CFRP with lightning protection metals and conductive or weakly conducting materials The device includes a sensor handheld module and a controller control and display module...

Page 9: ...lightning protection metals Power supply NiMh battery Temperature Operation 10 40 C 50 104 F Storage 0 45 C 32 113 F Dynamic tilt prevention ON OFF Parameters Statistical functions with recording and...

Page 10: ...on foils 1 Operating instructions and test certificate 1 Carry case Carefully remove the components of the measuring system from the packaging and ensure that the goods are forwarded in such a way tha...

Page 11: ...Page 11 Initial Operation FSC1 7 FSC1000 4 Initial Operation 4 1 Dimensional Drawings 68 2 68 123 4 84 48 1 89 54 2 13 Fig 3 Sensor dimensions in mm inches Active measuring surface sensor...

Page 12: ...Page 12 Initial Operation FSC1 7 FSC1000 128 5 04 43 1 96 230 9 06 238 9 37 Fig 4 Controller dimensions in mm inches...

Page 13: ...e battery Connect the battery to the charger The LED of the charger signalizes the charging process Fig 5 Controller with battery The state of charge of the battery is displayed on the controller scre...

Page 14: ...es sufficient time to adapt to the ambient temperature LED On USB Sensor On button 5 2 General Controller Operation The controller can be operated using the touchscreen or the keyboard The touchscreen...

Page 15: ...sensor is established when the sensor s serial number is displayed at the bottom of the screen see Fig 8 The device is ready for operation If a connection to the sensor cannot be established this is...

Page 16: ...ment by pressing keys 1 2 or 3 on the controller Both Start measurement by pressing the button on the sensor or keys 1 to 3 Measure count 1 2 3 4 5 All three parameters ensure repeatable measurement r...

Page 17: ...Page 17 FSC1 7 and FSC1000 Operation FSC1 7 FSC1000 Gray highlighted fields allow for a selection Fig 10 Setup menu...

Page 18: ...t table Manage Calibrations Manage delete existing adjustment tables New Calibration Create new adjustment table The measuring system can be calibrated on various sub strate materials Potential substr...

Page 19: ...Fig 12 Data set selection Fig 13 Data set details Select the desired file and confirm with the Load button Once an adjustment is loaded a window will appear with the basic information regarding the a...

Page 20: ...Managing Adjustment Tables You can manage existing data sets in the Manage Calibration submenu Possible options Del Deletes the selected data set Load Loads the selected data set ESC Back to measurin...

Page 21: ...r recertify the foils after the fact Switch to the New Calibration menu The adjustment can be started when Ready is displayed in the status window Fig 15 Adjustment ready You must select at least two...

Page 22: ...ge of max 100 to 150 m over the entire measuring range Make sure that the measuring area of the sensor the calibration foil and the measur ing point are clean Place the second calibration foil on the...

Page 23: ...asure count Measure delay and Max delta in the setup these settings will also be used for the adjustment i To carry out the adjustment set the Max delta parameter to the smallest possible value and se...

Page 24: ...List of support points during an adjustment You can change a selected support point using the Edit button Use the ESC button to cancel an adjustment series the system will return to measuring mode Us...

Page 25: ...e deviation delete the support point and recreate it see Fig 19 Fig 21 Test result Use the Save Calibration Data field to assign a name to the adjustment data set Then press Save Fig 22 Name assignmen...

Page 26: ...s the button on the sensor or keys 1 to 3 on the controller The measurement will start 300 ms after you release the measuring button Fig 23 Measurement preparation Fig 24 Display of measurement result...

Page 27: ...sted measuring range If the individual measurements have a deviation that is greater than the default Setup Max Delta menu the measurement was not successful The system will generate the message Diff...

Page 28: ...alyze the measurement data the mean value and the standard deviation are determined for the measured values within the values listed Switch to the Mean Measurement menu on the start screen Last Curren...

Page 29: ...cted and removed from the list in the log file but they are not used however for the statistical data set The measured values displayed in the statistic and overview are used for the statistics functi...

Page 30: ...ta menu on the screen Fig 31 Measurement file selection 2018_07_10_000 Mean 104 m SD 18 19 m Count 3 Min 90 m Max 130 m 12 16 49 129 m 12 16 05 131 m 12 17 38 130 m 12 17 46 93 m 12 17 54 90 m up down...

Page 31: ...ted file Load Opens the selected file with the data viewer see Chap 5 8 2 Esc Returns to the start screen Fig 33 Managing the measuring files 5 8 4 Saving Exporting Measurement Data You can use the Sa...

Page 32: ...ion FSC1 7 FSC1000 After the data has been successfully saved this is displayed Fig 34 Saving the measurement data If no data carrier is connected or it is not recognized you will see a message Fig 35...

Page 33: ...erial defects is 12 months from delivery Within this period defective parts except for wearing parts will be repaired or replaced free of charge if the device is returned to MICRO EPSILON with shippin...

Page 34: ...lt cannot be clearly determined always send the complete measuring system to MICRO EPSILON MESSTECHNIK GmbH Co KG K nigbacher Strasse 15 94496 Ortenburg Germany Tel 49 0 8542 168 0 Fax 49 0 8542 168 9...

Page 35: ...Ver 2 Ver 2 1 Ver 3 Digia Qt LGPL Exception version 1 1 Raspbian Jessie Raspbian Kernel 4 4 34 v7 GNU General Public License GPL 2 GNU GPL plus the GCC Runtime Library Exception ICU License ICU 1 8 1...

Page 36: ...N MESSTECHNIK GmbH Co KG K nigbacher Str 15 94496 Ortenburg Germany Tel 49 0 8542 168 0 Fax 49 0 8542 168 90 info micro epsilon com www micro epsilon com X9751382 A051109SWE X975382 A05 MICRO EPSILON...