Commands
colorCONTROL MFA-5
6.2.6
Baud Rate
6.2.6.1 Set Baud Rate
setbaudratexxxxxx
Command
Description
Received
Example
Note
setbaudratexxxxxx
Set baud rate
Default: 57600
OK
setbaudrate019200
Output format: OK
9600, 19200,
38400, 57600,
115200, 230400
Only applicable for
the colorCONTROL
MFA-5 <-> PC
connection
i
The baud rate between the colorCONTROL MFA-5 and MFA-5-M is 115200 and cannot be changed.
The baud rate between the colorCONTROL MFA-5 and the PC can be set individually.
6.2.7
Checkpoint Capture Example
Example: capture215 3
capture = start test
2
= test time -> 200 ms, MED
1
= sensitivity -> High, 9x9 color chip matrix
5
= sensor -> 5. Sensor
Space character
3
= MFA-5/-M number 3 -> 3. MFA-5 in series
A carriage return CR (0x0d) must be sent after every command.
Every received character string is terminated with CR
i
A space character must be sent between the command and the number of the colorCONTROL MFA-5.
If you use several colorCONTROL MFA-5, the command
testcon
must be sent before the test value
.