Promet I.S. User’s Manual
Michell Instruments
27
INSTALLATION
2.4.3.2 Optional Pressure Transmitter Wiring
HAZARDOUS AREA INFORMATION
The pressure transmitter is certi
fi
ed intrinsically safe for use
in hazardous areas by Baseefa and is covered by EC-type
examination certi
fi
cate BAS01ATEX1125X. The instrument
conforms to the standards speci
fi
ed in EN 50014:1997 + Amds 1
& 2, EN 50020:1994 and EN 50284:1999 with certi
fi
cation code
Ex ia IIB T4 (-20°C
≤
Ta
≤
75°C).
Before using the pressure transmitter in any hazardous
environment ensure that personnel are completely familiar
with the above standards relating to the certi
fi
cation of
this instrument and also with the intrinsically safe system
information in the BAS01ATEX1125X certi
fi
cate or equivalent
codes of practice in the country of installation.
If the pressure transmitter is not ordered together with the
analyzer, it is the user’s responsibility to make sure that the
pressure transmitter is compatible with the I.S. barrier in the
control unit.
Refer to ATEX/CSA/IECEx certi
fi
cates for the dew-point and
optional pressure transmitters’ connection cable requirements
which stipulate maximum permissible mutual capacitance and
inductance to resistance ratio.
NOTE: A pressure transmitter will only be provided if the model with active
pressure compensation (order code:
P1/P2
) is ordered. For the model without
active pressure compensation (order code:
G
), skip this section.
a.
The cable connection to the pressure transmitter is made via the removable
connector. Loosen the central screw (1) to remove the connector from the
transmitter.
Figure 19
Pressure Transmitter Connector
Summary of Contents for Promet I.S.
Page 53: ...Promet I S User s Manual Michell Instruments 45 APPENDIX A Appendix A Technical Specifications...
Page 61: ...Promet I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Modbus RTU Comms...
Page 63: ...Promet I S User s Manual Michell Instruments 55 APPENDIX C...
Page 67: ...Promet I S User s Manual Michell Instruments 59 APPENDIX F Appendix E System Drawings...
Page 69: ...Promet I S User s Manual Michell Instruments 61 APPENDIX F E 2 FM Approved System Drawing...
Page 75: ...Promet I S User s Manual Michell Instruments 67 NOTES...
Page 76: ......
Page 77: ......
Page 78: ...http www michell com...