Promet I.S. User’s Manual
10
97221 Issue 5, March 2017
INSTALLATION
2.2 Operating
Requirements
2.2.1 Environmental
Requirements
The Promet I.S. sensor assembly is intrinsically safe and designed to be installed on-
site, indoors or outdoors, directly at the point of measurement within a Hazardous
Area. The sensor assembly is ATEX and IECEx or CSA and FM certi
fi
ed (to be speci
fi
ed
at time of order). To operate correctly, the sensor assembly must be installed within
a suitable sampling system (Michell Instruments can supply standard and custom
designed sampling systems).
The Promet I.S. control unit is NOT designed for use in a Hazardous Area and should
only be installed in a safe area. The control unit is intended for indoor installation only
and operates within environmental limits of 0 to +50°C (+32 to +122°F) and <90%RH.
The control unit contains built-in isolation barriers permitting connection, direct from
the Hazardous Area, of the Promet I.S. sensor assembly.
2.2.2 Power
Requirement
The Promet I.S. control unit can be ordered for either of the following electrical supplies:
•
85 TO 265 V AC, 47/63 Hz @ 4W max 10W
OR
•
10 to 72 V DC @ 4W max 10W
The Promet I.S. sensor assembly is powered through the 2-wire signal cable directly
from the control unit.
Please refer to the Promet I.S. Premium Sampling System Instructions for the power
requirement of the sampling system.
2.3 Mounting
2.3.1 Control Unit Installation
The Promet I.S. control unit is contained in a 19” sub-rack case (size 3U). It should
be installed in a 19” rack using the mounting holes provided. It should be placed in a
position free from any appreciable vibration and shaded from direct sunlight.
NOTE: The materials and construction of the control unit allow for operation
in an indoor, clean, non-hazardous only, control room environment.
CHANNEL 1
CHANNEL 2
CHANNEL 3
CHANNEL 4
483mm
(19”)
375mm + 100mm for cable clearance
(14.75” + 3.9”)
3U
132mm
(5.2”)
Figure 5
Dimensions of the Control Unit
Summary of Contents for Promet I.S.
Page 53: ...Promet I S User s Manual Michell Instruments 45 APPENDIX A Appendix A Technical Specifications...
Page 61: ...Promet I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Modbus RTU Comms...
Page 63: ...Promet I S User s Manual Michell Instruments 55 APPENDIX C...
Page 67: ...Promet I S User s Manual Michell Instruments 59 APPENDIX F Appendix E System Drawings...
Page 69: ...Promet I S User s Manual Michell Instruments 61 APPENDIX F E 2 FM Approved System Drawing...
Page 75: ...Promet I S User s Manual Michell Instruments 67 NOTES...
Page 76: ......
Page 77: ......
Page 78: ...http www michell com...