background image

Page 4 of 8

D  MODES

"Day" or "Delayed" modes

The next three advancing settings are the D5D10 and D12.  These are the D for "Day" or "Delayed" modes.
The D5 setting will cause the EDS-D1 unit to delay responding to breathing actions until it senses a pressure
altitude
 of 5,000 ft. and above. The D10 setting operates respectively.  The D12 setting is calibrated to start
operation at a pressure altitude of
about 11,000 ft. ±250 ft.  This lower
trip-point setting is intentional so the
unit will comply per FAR part
23.1447 throughout a wide range of
barometric pressures. For example,
because the EDS-D1 unit has no user
adjustments for changes in
barometric pressure, it operates
simply from pressure altitudes such
as your body does. Therefore, if the
unit is being used while the
barometric pressure is low, it will
start operation sooner (at a lower
MSL flight altitude) than it will if the
barometric pressure is high. Therefore, the EDS-D1 is calibrated to trip a bit low to keep the D12 setting from
never violating the current FAR 23.1447 for oxygen use at 12,500 ft. MSL. The D5 is calibrated for a pressure
altitude
 of 5,000 ft. and  D10 is calibrated for a pressure altitude of 10,000 ft.

3 0

2 5

2 0

1 5

1 0

  5

  0

PRESS ALT  X  1K

PULSE  UNITS

D 5

D 1 0

D 1 2

D Modes

Unit delays 
activation 
until at and 

above 
selected 
pressure

altitudes 

D12

Switch as seen in one of the 

"Day" or "Delay" settings.

N  MODE

"Night" or "Now" mode

The first power on setting advanced from off is the N for "Night" or "Now" setting.  At this setting the EDS-D1
will respond to breathing actions at all altitudes with the standard delivery protocol.  The EDS-D1 will deliver
pulses of oxygen at the effective flow rate
of 1.0 liter/min. per 10,000 ft. for pressure
altitudes
 up to about 34,000 ft.

NOTICE

Using any of the on settings will result in
the correct (or more) amount of oxygen for
any given altitude. None of the on settings
will inadvertently result in inadequate
amounts of oxygen for any given altitude
in which the EDS operates.

N

Switch as seen set to

"Night" or "Now" mode

3 0

2 5

2 0

1 5

1 0

  5

  0

PRESS ALT  X  1K

PULSE  UNITS

N mode

Unit active at all 
pressure altitudes
delivering more 
with increasing 

altitude.

A

B

C

A

B

C

Align tube

Push

Push tube 

f i r m l y

Push 

f i r m l y

Push in 

collets

Pull tube
out while

pushing

collets

in about

1 cm.

To remove the polyurethane service
line you must push in the connector
collar while you pull on the
polyurethane line.  Do not pull the
polyurethane line without pushing in
the collar.  It will damage the
connector rendering it useless.

Summary of Contents for EDS-D1

Page 1: ...evice is intended for supplemental breathing while exposed to high altitudes On demand and up to 25 BPM this device delivers a pulse of supplied oxygen at the start of each inhalation phase Higher alt...

Page 2: ...n High E S Company assumes no responsibly for any accidents injury or death that may result from the misuse of this device equipment This includes any usage of this device equipment outside the scope...

Page 3: ...g tests are showing that 90 of the oxygen the EDS D1 provides to the lungs goes directly into the blood The EDS D1 synchronous inhalation pulse demand technique is currently the most efficient way kno...

Page 4: ...hing in while pushing door outwards fig 1 This is best done if the unit is held with the battery door facing up at you and out The EDS D1 unit uses a standard 9 nine volt alkaline battery connector sy...

Page 5: ...Positive GRN RED indicators for FLOW NO FLOW oxygen delivery status Provides Apnea alarm to inform user of kinked pinched disconnected lines or holding breath too long Saves existing oxygen supply by...

Page 6: ...10 000 ft 3 0 2 5 2 0 1 5 1 0 5 0 PRESS ALT X 1K PULSE UNITS D 5 D 1 0 D 1 2 D Modes Unit delays activation until at and above selected pressure altitudes D12 Switch as seen in one of the Day or Delay...

Page 7: ...r for use with regulators that are flow rate adjustable F MODES Floor Face Mask These next four settings deserve your attention The F Mode settings are F10 F15 F20 and F25 called the floor or facemask...

Page 8: ...negative and the outside ring is positive A reverse polarity diode 1 A in the EDS unit helps prevent damage from accidental short term reverse voltage occurrences Misc notes The EDS model D1 has a bas...

Page 9: ...itially off turn on the cylinder valve slowly to allow gas from the regulator to back charge the outlet side of the EDS D1 unit Next turn the EDS D1 unit on During the time the EDS D1 unit initializes...

Page 10: ...r normal operating conditions Storage Approx 4 6 months while in the off setting battery drain is about 1 a in the off setting Make sure that the battery is removed from the unit during long term stor...

Page 11: ...r for use with an EDS unit NOTICE IMPORTANT EDS INLET PRESSURE NOTICE If The EDS D1 unit s will not be used with the XCR XCP micro lite system or EDS STR regulator the alternate regulator must be able...

Page 12: ...s and capability never offered before Complete single and multi place carry on EDS system packages are available along with padded fabric carry strap down harnesses for most of our kevlar fiber and al...

Reviews: