background image

Color-coded 6 mm.

BLUE

outlet Quick-Connect union

connects to the cannula or

face mask

Color-coded 4 mm.

RED

inlet Quick-Connect

union connects to

the primary reducing

regulator

Power On/Off

and Control

Switch

Operational and

precautionary

front face-place

Audio port.

Sound from

flow-fault

and

apnea

conditions only

POSITIVE

FLOW

!

FLOW FAULT

APNEA

LOW BATT

STATIC PORTS

DO NOT COVER

EDS MODEL D1

MH

INLET TO

SUPPLY / CYLINDER

OUTLET TO

CANNULA / FACE MASK

Green LED

Red LED

Static ports

DO NOT COVER

EDS-D1

Page 2 of 8

High brightness green

LED for indication of

positive flow of

oxygen for each

inspiration effort.

High brightness

red LED for

indication of

oxygen

flow-faults,

apnea and low

battery condition.

External Battery

Cannula

Inside (tip) is neg (-)

Outside (ring) is pos (+)

EDS Unit

EPS

11 - 28 VDC input

The EDS-EPS is a floating 
ground unit suitable for 
negative or positive ground 
systems.  Be sure to ground it 
to the main system ground.

System Ground

1

2

3

9 Volt

A l k a l i n e

Replacing the battery in the EDS unit

Remove the battery door by holding the unit with both hands
and pressing in with your thumbs at the point of the arrow
pushing in while pushing door outwards (fig 1.)  This is best
done if the unit is held with the battery door facing up at you
and out.  The EDS-D1 unit uses a standard 9 (nine) volt
alkaline battery connector system.  This is a very positive-
connecting type connector that requires a bit more force to
connect and disconnect to the battery.  This is a superior type
connector for equipment subjected to rough out-door
treatment.  Therefore, please take extra care in removing and
replacing the battery making sure not to pull on the wires.
Use only alkaline type batteries with the EDS-D1 unit.

If the EDS -D1 unit will be

operating outside in very cold

temperatures, premature

battery depletion may be

likely. An external battery

connector can be wired up to
use an external 9 volt battery

placed somewhere warmer

such as your flight suit. It

can also be used as an

emergency backup,

electronically replacing the

internal battery with the

external battery that is

plugged in the external jack..

Up to (8) eight  EDS-D1 units can
be powered from our optional EPS
(External Power Supply) unit if it is
desirable to use the power system
from an aircraft. The EPS is a 10
volt @ 500 ma. low-drop-out
voltage regulator and filter unit that
operates from 11 to 28 volts. It
filters out glitches caused by
hydraulic pumps, strobe lights and
other high-energy devices. It works

Summary of Contents for EDS-D1

Page 1: ...evice is intended for supplemental breathing while exposed to high altitudes On demand and up to 25 BPM this device delivers a pulse of supplied oxygen at the start of each inhalation phase Higher alt...

Page 2: ...n High E S Company assumes no responsibly for any accidents injury or death that may result from the misuse of this device equipment This includes any usage of this device equipment outside the scope...

Page 3: ...g tests are showing that 90 of the oxygen the EDS D1 provides to the lungs goes directly into the blood The EDS D1 synchronous inhalation pulse demand technique is currently the most efficient way kno...

Page 4: ...hing in while pushing door outwards fig 1 This is best done if the unit is held with the battery door facing up at you and out The EDS D1 unit uses a standard 9 nine volt alkaline battery connector sy...

Page 5: ...Positive GRN RED indicators for FLOW NO FLOW oxygen delivery status Provides Apnea alarm to inform user of kinked pinched disconnected lines or holding breath too long Saves existing oxygen supply by...

Page 6: ...10 000 ft 3 0 2 5 2 0 1 5 1 0 5 0 PRESS ALT X 1K PULSE UNITS D 5 D 1 0 D 1 2 D Modes Unit delays activation until at and above selected pressure altitudes D12 Switch as seen in one of the Day or Delay...

Page 7: ...r for use with regulators that are flow rate adjustable F MODES Floor Face Mask These next four settings deserve your attention The F Mode settings are F10 F15 F20 and F25 called the floor or facemask...

Page 8: ...negative and the outside ring is positive A reverse polarity diode 1 A in the EDS unit helps prevent damage from accidental short term reverse voltage occurrences Misc notes The EDS model D1 has a bas...

Page 9: ...itially off turn on the cylinder valve slowly to allow gas from the regulator to back charge the outlet side of the EDS D1 unit Next turn the EDS D1 unit on During the time the EDS D1 unit initializes...

Page 10: ...r normal operating conditions Storage Approx 4 6 months while in the off setting battery drain is about 1 a in the off setting Make sure that the battery is removed from the unit during long term stor...

Page 11: ...r for use with an EDS unit NOTICE IMPORTANT EDS INLET PRESSURE NOTICE If The EDS D1 unit s will not be used with the XCR XCP micro lite system or EDS STR regulator the alternate regulator must be able...

Page 12: ...s and capability never offered before Complete single and multi place carry on EDS system packages are available along with padded fabric carry strap down harnesses for most of our kevlar fiber and al...

Reviews: