6
MGF operating instruction manual
Canister is equipped of 3 probes ( Fig.4 - 3) that control liquid residual liquid level inside the
canister: if liquid residual level exceed the shorter probe, draining pump automatically starts
and empties the canister. This operation takes a few seconds and dentist may feel a short
suction reduction in the manifold.
In case of lower residual liquid level, the canister will empty at the end of suction unit
operating and in this case the drainage is made by gravity without draining pump only with
drainage valve opening (Fig.4 - 2).
During surgical operations foams made by blood and the turbulence of aspirated air may
affect the probes function and stop the aspiration: for this reason is recommended the use of
solid antifoaming agent.
Canister is also equipped with a solid residuals filter (Fig.4 - 5). Solid residuals may be
aspirated during standard operating and in case they reach aspiration or draining pump may
damage your Silent unit. Solid residuals filter prevents this problem.
In case of block of the suction unit due to block of the cooling system, breakdowns or other
problem immediately contact your MGF dealer.
Fig. 4
1. Canister tank
2. Canister body
3. Draining valve
4. Probes
5. Solid residuals filter
6. Canister air inlet
7. Canister air outlet