5 Troubleshooting
page
19
ELC N6_N10-V3.1-09/11-GB subject to technical alterations
5 Troubleshooting
Troubles Possible
causes
Corrective
action
No RELEASE after having
switched on the ELC.
Relay RELEASE open
The ELC is not provided
with operating voltage
Check fuses and voltage of
terminals
Lamp is switched-ON
(control input START at
HIGH)
Switch-OFF lamp
Fault in ELC
Contact manufacturer,
see 5.1
Lamp is too hot
Let lamp cool down, increased
cooling probably required
Lamp does not fire after being
switched on, and the relay
RELEASE opens
Lamp not connected
Check plug contacts between
ELC and lamp
Lamp defective
Replace lamp
5.1 Repair
ELC
Only the manufacturer is permitted to repair the ELC. In case of failure please contact the
following address: