background image

 

 

 
 
 

Sys*Stim

®

 206 

Instruction Manual 

 

 

 
 
 
 
 

 

 

 

 

 
 

 

 

 

 

 

®

 

1333 South Claudina Street  •  Anaheim, CA 92805, U. S. A. 

Toll Free: (800) 854–9305 • Telephone: 1(714) 533–2221 • FAX: 1(714) 635–7539 

Web Site: 

www.mettlerelectronics.com

  •  Email: 

[email protected] 

IR6–05   

 

 

 

 

 

        Rev. E_4/4/13 

Copyright ©  1986, 2004 by Mettler Electronics Corp.—Anaheim, CA 

Summary of Contents for SYS-STIM 206

Page 1: ...ina Street Anaheim CA 92805 U S A Toll Free 800 854 9305 Telephone 1 714 533 2221 FAX 1 714 635 7539 Web Site www mettlerelectronics com Email mail mettlerelectronics com IR6 05 Rev E_4 4 13 Copyright 1986 2004 by Mettler Electronics Corp Anaheim CA ...

Page 2: ...Mettler Electronics Corp Rev E_4 4 13 2 ...

Page 3: ...ions 13 4 Operating Instructions 15 4 1 A Note About Electrodes 16 4 2 General Operating Instructions 16 4 3 Operating Instructions 17 5 Indications Contraindications Precautions and Adverse Reactions 19 5 1 Indications 19 5 2 Contraindications 19 5 3 Warnings 20 5 4 Precautions 21 5 5 Side Effects Adverse Reactions 22 6 Maintenance and Troubleshooting 23 6 1 Cleaning the Sys Stim 206 23 6 2 Routi...

Page 4: ...ronics Corp Rev E_4 4 13 4 List of Figures No Title Page 1 1 Sys Stim 206 5 4 1 Front membrane panel and LED indicators 15 7 1 Narrow Pulse Illustration 25 7 2 Wide Pulse Illustration 26 7 3 AC Pulse Illustration 26 ...

Page 5: ...alanced with zero net charge There is a surge mode set at 6 seconds on and 18 seconds off at either 35 or 83 pulses per second The Sys Stim 206 is portable and can be combined with any Sonicator therapeutic ultrasound unit for combination therapy Figure 1 1 Sys Stim 206 1 2 Introduction to This Manual Read the contents of this manual before treating patients with the Sys Stim 206 This manual has b...

Page 6: ...service facility authorized by Mettler Electronics to perform warranty repair work A service manual for the Sys Stim 206 is available from Mettler Electronics Corp for a nominal charge 1 4 Caution Federal law restricts the sale of this device to or on the order of a physician dentist veterinarian or any other practitioner licensed by law of the state in which he practices Use of controls or adjust...

Page 7: ...rranted against defects in materials and workmanship for a period of one year from date of purchase During the applicable warranty period Mettler Electronics Corp will at its discretion either repair or replace the Product without charge for these types of defects For service under this warranty the Product must be returned by the buyer within the applicable warranty period to Mettler Electronics ...

Page 8: ...ced within one year from the date any claim arises and must be brought only in the state or federal courts located in Orange County California Some states do not allow limitations on how long an implied warranty lasts or the exclusion or limitation of incidental or consequential damages so the above limitations or exclusions may not apply to the buyer This warranty gives the buyer specific legal r...

Page 9: ...econds and restarts at 0 00 after each 29 minute time period The Go button starts timer and begins the treatment by providing the AC pulse or DC energy to the electrode jack output if the intensity control is in fully counterclockwise position Reset at the time the Go switch is depressed The Hold button halts the timer without disturbing the remaining treatment time and stops the electrical output...

Page 10: ... the Wide Pulse button to enter the Wide Pulse mode The green indicator to its left illuminates when activated to show that the Wide pulse is selected Its shape is the same as the Narrow pulse shape except that the pulse duration is 300 μs of one polarity followed by amplitude for 1200 μs of reversed polarity Press the AC Pulse button to enter the AC Pulse mode The green indicator to its left illu...

Page 11: ...ring the initial 15 second period described above The green indicator next to Pulse Frequency control is on during that period of time The pulse frequency in the Surge mode is displayed in the meter when Surge is selected and output is inactive Hold mode Two frequencies are available 35Hz or 83 Hz by depressing the button of the currently selected pulse waveform Each successive depression of the s...

Page 12: ... per second In the fully clockwise position the pulse frequency is approximately 83 pulses per second Frequency increases with a clockwise rotation and decreases with a counterclockwise rotation A green indicator located under the word pulse near the pulse frequency control is illuminated when the pulse energy is present at the electrode jack output O Mains Off Attention consult instruction manual...

Page 13: ...prong plug This plug provides grounding for the Sys Stim 206 Do not defeat its purpose by using 3 to 2 prong adapters or any other means of attaching to a wall outlet 3 Plug the electrode cable ME 2260 into the electrode jack 4 The Sys Stim 206 may be susceptible to interference originating from shortwave diathermy units operating in close proximity to it Avoid operating the Sys Stim 206 adjacent ...

Page 14: ... until Hold is pressed or treatment time expires Polarity Polarity is reversible by depressing the Reverse Polarity switch Narrow Pulse Pulse frequency Peak Volts Same as AC Small muscles or muscle groups or pain management Surge Same as AC except Narrow Pulse is surged Polarity Same as AC Wide Pulse Pulse frequency Peak Volts Same as AC Medium muscles or muscle groups Surge Same as AC except Wide...

Page 15: ...Sys Stim 206 Instruction Manual Rev E_4 4 13 15 Section 4 Operating Instructions 00 00 000 Figure 4 1 Front membrane panel and LED indicators ...

Page 16: ...e to high current density do not use electrodes smaller in surface area than 2 in diameter 7 Do not use conductive carbon electrodes with this product especially in the DC Continuous mode 8 Most self adhesive electrodes are inappropriate for use with DC Continuous stimulation and may be permanently rendered inoperable if DC Continuous is used with them 9 To insure efficient current conduction nece...

Page 17: ... guide at the end of this manual Narrow pulse green indicator lit Timer indicator decimal point only Output meter 0 Peak volts green indicator lit 5 Place electrode pads over area to be treated using proper technique Note To reduce skin to pad resistance strap the sponge electrode firmly to the skin without impairing circulation Note No straps are necessary when using self adhesive electrodes 6 En...

Page 18: ...de and option combination 11 Rotate the Intensity Control clockwise to increase output to the desired treatment level Adjust intensity in Surge mode only when the green Pulse Frequency indicator shows pulses present during the first 15 seconds of initial Surge mode operation After that you will only be able to adjust the intensity level up during the On time of the Surge cycle 12 Rotate the Pulse ...

Page 19: ...vention or retardation of disuse atrophy 6 Muscle re education 7 Maintaining or increasing range of motion 8 Denervated or partially denervated muscles may be stimulated using the pencil electrode to turn on and off the stimulus 5 2 Contraindications 1 Application of galvanic DC current can produce skin irritation leading to possible damage Knowledge of physiological effects of continuous direct c...

Page 20: ...mouth may cause severe spasm of the laryngeal or pharyngeal muscles These contractions may be strong enough to close the airway or cause difficulty in breathing 14 Do not apply stimulation for undiagnosed pain syndromes until etiology is established 15 Do not apply electrodes directly over the eyes or inside body cavities 16 Do not use electrical stimulation in conjunction with high frequency surg...

Page 21: ...wing long term application 5 Avoid placing electrodes directly over open wounds since current density tends to concentrate in these areas 6 Use extreme caution when treating desensitized areas or on patients who may not be able to report discomfort or pain 7 Use caution in applying electrical stimulation over areas where there is a loss of normal skin sensation 8 Adequate precautions should be tak...

Page 22: ...rs adequate directions for use of the device with the drug 16 Turn on the Sys Stim 206 before applying electrodes to the patient 5 5 Side Effects Adverse Reactions 1 Skin irritation and burns beneath the electrodes have been reported with the use of electrical muscle stimulators 2 Possible allergic reactions to tape gel or electrodes may occur ...

Page 23: ...outine cleaning of the sponge electrodes and electrode cables use soap and water and rinse thoroughly Dry cables and their connections after cleaning 6 2 Routine Maintenance 1 Standard medical electrical safety checks should be performed annually by qualified biomedical engineers or technicians trained to perform these procedures 2 Inspect electrode cable and associated connector for damage 6 3 Tr...

Page 24: ...at the Reset position fully counterclockwise Note 2 The Sys stim 206 is equipped with extended high current sensors that will shut down the output immediately upon an error condition for the protection of the patient The extended high current condition may be reset by turning the Main power control off for at least 3 seconds before turning it back on Caution An extended high current condition that...

Page 25: ... net DC Narrow and Wide pulses Symmetrical biphasic with zero net DC AC pulse Continuous Direct Current Voltage 0 TO 102 VDC Current 30 mA max DC Continuous 100 mA peak Narrow Pulse Wide and AC pulses Phase Duration Narrow Pulse Wide Pulse AC 100 μs followed by 400 μs at amplitude 300 μs followed by 1200 μs at amplitude 300 μs in each direction Frequency Pulse Mode 1 83 Hz Surge Cycle 6 seconds On...

Page 26: ...Mettler Electronics Corp Rev E_4 4 13 26 20 Volts Division 0 2ms Division Figure 7 2 Wide Pulse Illustration 20 Volts Division 0 2 ms Division Figure 7 3 AC Pulse Illustration ...

Page 27: ... electrode 3 5 x 7 2005 12 Sponge inserts 3 5 x 7 2006 1 Sponge electrode 8 x 10 2007 12 Sponge inserts 8 x 10 2008 4 Electrode straps 24 2009 4 Electrode straps 48 2023 Pencil electrode set with push button stimulation control includes handle 4 different sizes of stainless steel spot electrode tips and carrying case 2027 Pin to banana adapter plug set to be used with ME 2026 2260 or 2201 electrod...

Page 28: ...trodes with lead wires case of ten packages four electrodes pkg 2703 V Trode 2 75 diameter round electrodes with lead wires case of 10 packages four electrodes pkg 2704 V Trode 2 x 4 oval electrodes with lead wires case of 10 packages four electrodes pkg 2705 V Trode 2 square electrodes with lead wires case of 10 packages four electrodes pkg ...