b. The following menu will be displayed:
c. Select the desired waveform by pressing
UP (F2)
or
DOWN (F3)
and then
Select (F4)
. Press
CANCEL (F5)
to
cancel the selection. After selection the main menu will
reappear.
d. Select
SENS. TEST
(F2)
. The following display will ap-
pear:
e. Upon completion of testing the results will be displayed un-
der RESULT. Press
SENS. TEST. CANCEL
(F5)
to can-
cel the test.
5.4 Refractory Period Test
1.
General.
This test is
used to test the time interval in millisec-
onds (ms) during which the pacemaker is insensitive to any ex-
ternal inputs. QA-45 does this by measuring the maximum time
interval after the generation of a pacer pulse, and maximum
time interval after a QRS wave.
a.
Refractory Period
. A time interval in milliseconds, during
which a pacemaker is insensitive to any external inputs. If a
QRS is detected during this period, the pacemaker ignores
it. On the other hand, if a QRS is detected outside the re-
fractory interval, then the pacemaker resets its internal
5
Summary of Contents for QA-45
Page 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Page 5: ...This page intentionally left blank v...
Page 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Page 26: ...This page intentionally left blank 9...
Page 28: ...2...
Page 38: ...4...
Page 42: ...This page intentionally left blank 8...
Page 58: ...This page intentionally left blank 10...
Page 60: ...A 2...
Page 61: ...A 3...
Page 62: ...A 4...
Page 63: ...A 5...
Page 64: ...A 6...
Page 65: ...A 7...
Page 66: ...A 8...
Page 67: ...A 9...
Page 68: ...A 10...
Page 69: ...A 11...
Page 70: ...A 12...
Page 71: ...A 13...