Choose a test variant by pressing
UP (F2)
or
DOWN (F3) or
TEST PULSE (F1)
. Press
CANCEL (F5)
to cancel selection.
For ‘ECG0’, ‘ECG+’ and ‘ECG-’ see Chapter 6, Control and
Calibration. For ‘A/D-read’, see paragraph 7.3.7, page 7-5.
Memory’ is for factory testing. Also, see paragraph 4.3.5, page
4-3.
11.
REMOTE CONTR. (Remote Control) (F4)
enables communi-
cation with a PC with test automation software. Required soft-
ware: PRO-Soft QA-40/45.
3.4 Menu and Messages:
Transcutaneous
Pacemaker Mode
1.
Startup Screen
. The following screen will be displayed for 2
seconds after the QA-45 has been switched on.
2.
Main Menu
a.
M
ain Menu Bar (Page 1) - Mode switch in PACE position.
Note
QA-40M has an internally gener-
ated test pulse. The control pulse
is set at 1.2 Joules in the Low
range and 28.5 Joules in the High
range. The test pulse is not a cali-
bration pulse, and should not be
used as an indication of the gen-
eral accuracy of the instrument.
The test pulse is a good control for
testing functions.
6
Summary of Contents for QA-45
Page 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Page 5: ...This page intentionally left blank v...
Page 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Page 26: ...This page intentionally left blank 9...
Page 28: ...2...
Page 38: ...4...
Page 42: ...This page intentionally left blank 8...
Page 58: ...This page intentionally left blank 10...
Page 60: ...A 2...
Page 61: ...A 3...
Page 62: ...A 4...
Page 63: ...A 5...
Page 64: ...A 6...
Page 65: ...A 7...
Page 66: ...A 8...
Page 67: ...A 9...
Page 68: ...A 10...
Page 69: ...A 11...
Page 70: ...A 12...
Page 71: ...A 13...