2. If PRO-Soft QA-40M/45 is being used, attach an RS-232 (null
modem/data transfer configured) cable to the 9-pin D-sub outlet
port located at the rear of the QA-45. Do not attach the printer
cable to the QA-45.
See below
. However, if you are not using
PRO-Soft QA-40M/45, and are sending directly to a printer for
printouts, attach the printer cable to the 25-pin outlet port.
2.3 Power
1.
Main On/Off Switch.
QA-45 should remain off for at least 5
seconds before switching on again, in order to allow the test cir-
cuits to discharge fully.
2.
Low Battery Power.
If battery power falls below 6.9 volts
(
±
0.3 volts), the display will show 'Change battery, and reset
system'. This means that the battery should either be replaced or
the instrument should be connected to a battery eliminator. The
NOTE
Some RS-232C cables are
missing the connection be-
tween the seventh and the
eighth wires in the cable. The
cable may still be called
NULL-modem, but it will not
work with the QA-45. Refer to
the PRO-Soft QA-40M/45
Users Manual for more informa-
tion.
2
Summary of Contents for QA-45
Page 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Page 5: ...This page intentionally left blank v...
Page 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Page 26: ...This page intentionally left blank 9...
Page 28: ...2...
Page 38: ...4...
Page 42: ...This page intentionally left blank 8...
Page 58: ...This page intentionally left blank 10...
Page 60: ...A 2...
Page 61: ...A 3...
Page 62: ...A 4...
Page 63: ...A 5...
Page 64: ...A 6...
Page 65: ...A 7...
Page 66: ...A 8...
Page 67: ...A 9...
Page 68: ...A 10...
Page 69: ...A 11...
Page 70: ...A 12...
Page 71: ...A 13...