Overview – OMNIS Sample Robot Pick&Place L
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
(front and rear), the titration head holder
can be moved downwards via the lift tower
so that the safety
encases the sample beaker.
A titration head that fits the sample beaker to be used can be utilized in
the titration head holder
. The titration head contains sensors, dosing
tips, cleaning accessories (and occasionally a rod stirrer or homogenizer)
that are required for the measurement.
Rear
Figure 8
Rear – Pick&Place module
1
Drain nozzle
2
Distributor
At the rear of the Pick&Place module there is a drain nozzle
to which
a tubing is connected via a tubing adapter. Any liquid that may escape will
be routed through this tubing into the waste canister. In case of an error,
this protects the Pick&Place module from damage.
that is used for connecting the rinsing tubing and
aspiration tubing is fitted on the back panel of the Pick&Place module.
Option for working with a homogenizer (Polytron PT 1300
D)
A homogenizer (Polytron PT 1300 D)
can be inserted in the titration
head
for applications which require homogenization of the sample.
A holder for homogenizers has to be mounted for guiding the homoge-
nizer
. For further information, see
Inserting and equipping the
titration head
.