![Metrohm Dialysis 6.5330.100 Instruction Manual Download Page 21](http://html1.mh-extra.com/html/metrohm/dialysis-6-5330-100/dialysis-6-5330-100_instruction-manual_1779676021.webp)
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC equipment Dialysis 6.5330.100
■■■■■■■■
15
■
Tighten another PTFE capillary, 0.5 mm ID / 1 m (6.1803.040), to
the outlet of the second pump tubing with black stoppers
(6.1826.340) using a pressure screw (6.2744.070).
Tighten the other end of the PTFE capillary to the waste container.
5 Connecting the inlet of the acceptor solution
■
Tighten a PTFE capillary (6.1803.040) to the inlet of the pump
tubing with orange/yellow stoppers (6.1826.320) using a pressure
screw (6.2744.070).
Immerse the other end of the PTFE capillary in the vessel contain-
ing the acceptor solution (6.1808.070 with 6.1602.150) and fas-
ten with a PVDF pressure screw (6.2744.000).
■
Tighten another PTFE capillary (6.1803.040) to the outlet of the
pump tubing with orange/yellow stoppers using a pressure screw
(6.2744.070).
Tighten the other end of the PTFE capillary to the acceptor inlet of
the dialysis cell
using a pressure screw (6.2744.000).
6 Connecting the outlet of the acceptor solution