Autolab/Electronic load combination
11 |
P a g e
Figure 6 – Connections overview without FRA module
1.
Connect the DAC164 output one on the front panel of the Autolab
(DAC164
1) to the DAC-1 input of the Dynload interface.
2.
Connect the Y-FRA/ADC output of the Dynload interface to the ADC164
input one on the front panel of the Autolab (ADC164
1).
3.
The DSG input of the Dynload interface must be shorted with the provided
50 Ohm termination plug (see Figure 7).
Figure 7 – A 50 ohm terminator plug must be used to short the DSG input when this input is
not used
DYNAMIC LOAD
External prog.
DAC-1
DSG
Y-FRA/ADC
I-M
15 V
+
-
50 Ω
Note
Make sure that the 50 Ohm plug is connected to the DSG input at all times!