4.3
Instrument problems
853 CO
2
Suppressor / Instructions for Use 8.853.8001EN
13
4.3 Instrument
problems
Problem
Cause
Remedy
LED "
Power
"
) does
not light up when in-
strument is switched
on.
•
Power supply
not con-
nected
•
6.2152.020 Power sup-
ply
faulty or
electronics
fault in 853 CO
2
Suppres-
sor
•
Connect the 853 CO
2
Suppres-
sor to the mains supply via the
external power supply (see
Section 2.2
•
Contact Metrohm Service.
Neither
LED "
Status
"
LED "
Ready
"
) lights up when in-
strument is switched on
(
LED "
Power
"
(
) lights
up).
•
LED
s faulty
•
Contact Metrohm Service.
Vacuum out of working
range.
LED "
Ready
"
does not light up after
instrument has been
switched on for at least
30 s.
•
Leak
in system (vacuum
too weak)
•
H
2
O Absorber Cartridge
blocked (vacuum too
strong).
•
Contact Metrohm Service.
•
Regenerate the
H
2
O Absorber
Cartridge
(see
Section 4.4.4
or 8.108.1046 Leaflet)
4.4 Care
and
maintenance
4.4.1 Care
The 853 CO
2
Suppressor requires adequate care. If it becomes excessi-
vely dirty then this could interfere with its functions and shorten the
working life of its robust mechanism and electronics.
Spilt chemicals and solvents must be wiped up immediately. The elec-
trical connections on the rear panel of the instrument (the mains con-
nection in particular) should be protected against contamination.
Although constructive measures to a large extent prevent liquid pene-
tration, if aggressive media should nevertheless penetrate the interior
of the instrument then the mains plug of the external 6.2152.020
Power supply should be pulled out immediately to prevent serious
damage to electronic components. If such damage should occur
please contact Metrohm Service.
The instrument should only be opened by specialists from Metrohm.
Summary of Contents for 853 CO2 Suppressor
Page 1: ...853 CO2 Suppressor Manual 8 853 8001EN ...
Page 2: ......