2 Installation
853 CO
2
Suppressor / Instructions for Use 8.853.8001EN
10
2.5 Connection
of
capillaries
The 853 CO
2
Suppressor is connected between the suppressor module
«MSM II» and the detector block.
1 Suppressor module «MSM II» – 853 CO
2
Suppressor
•
Connect the "Suppressor outlet capillary for eluent" (marked
with "Detector") to the
"
Eluent in
"
inlet (
) of the 853 CO
2
Suppressor using a 6.2744.010 Pressure screw.
2 853 CO
2
Suppressor – Detector block
•
Connect the "Inlet capillary to detector block" to the
"
Eluent
out
"
outlet (
) of the 853 CO
2
Suppressor using a 6.2744.010
Pressure screw.
2.6 Cartridges
Effective CO
2
removal requires that the air pumped through the de-
gassing cell should contain as little CO
2
as possible. This is achieved
by drawing the air through a 6.2837.000
CO
2
Absorber Cartridge
.
The
CO
2
Absorber Cartridge
can become blocked by moisture. This
can be prevented by connecting a 6.2837.010
H
2
O Absorber Car-
tridge
in front of it.
The cartridges are installed as follows:
1 Remove caps
•
Remove the sealing caps at the inlet and outlet of each
cartridge.
2 Connect CO
2
Absorber Cartridge
•
Insert the CO
2
Absorber Cartridge (3-layer filling, blue-brown-
3 Connect H
2
O Cartridge – CO
2
Cartridge
•
Insert the 6.1808.190 Adapter in the CO
2
Absorber Cartridge.
•
Attach the 6.1801.140 PVC tubing to the 6.1808.190 Adapter.
•
Insert the H
2
O Absorber Cartridge (filled with orange drying
beads) into the 6.1801.140 PVC tubing.
4 Attach the cartridges
•
Attach the two cartridges to the left-hand mounting support of
the 861 Advanced Compact IC or 820 IC Separation Center
using two 6.2027.070 Holders.