![Metrohm 793 IC Instructions For Use Manual Download Page 38](http://html1.mh-extra.com/html/metrohm/793-ic/793-ic_instructions-for-use-manual_1779567038.webp)
4 Malfunctions - Maintenance
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
34
40
40
40
40
39
39
39
39
38
38
38
38
37
37
37
37
Fig. 9: Assembling the actuator
37
37
37
37
Screw nut
39
39
39
39
Actuator rotor (6.2835.000)
38
38
38
38
Connection piece (6.2835.010)
with input and output leads
40
40
40
40
Actuator holder
6 Clean connection piece
x
Clean the sealing surface of connection piece
38
38
38
38
with the aid
of a lint-free cloth and ethanol.
7 Insert connection piece
x
Insert connection piece in actuator holder
40
40
40
40
in such a way
that connection "1" is at the top and that the three lugs on the
connection piece fit in the corresponding openings of the
holder.
x
Screw nut
37
37
37
37
onto the thread of actuator holder manually (do
not use tools).
8 Connect and condition the SP Module A
x
Reconnect the SP Module A to the IC system (see
section
2.3.2 and section 2.4.2
).
x
Before switching the actuator to the next position for the first
time rinse all 3 actuator units with solution for 5 min.