![Metrohm 736 GP Titrino Instructions For Use Manual Download Page 9](http://html1.mh-extra.com/html/metrohm/736-gp-titrino/736-gp-titrino_instructions-for-use-manual_1779554009.webp)
1.3 Storing methods
736 GP Titrino: Getting started
7
1.3 Storing methods
You learn how to operate the method memories.
The Titrino offers 2 types of method memories:
•
Internal method memory, key <user meth>
•
External method memory on the card, key <card>
Storing a method in the internal method memory
Storing a method in the internal method memory
2 x <user meth>
<enter>
<
<
<
>
<
<
<
> + <
>
>
>
<enter>
:
<QUIT>
<enter>
Press key <user meth> repeatedly until ">store method" ap-
pears in the display.
Enter an identification for your method, e.g. HCl:
Press key <
<
<
>. The display shows a row of letters. The first
position of this row is blinking. This is the input position.
You can move this row of letters pressing the <
<
<
> and
<
>
>
>. Press <
>
>
> until an "H" appears in the blinking posi-
tion and confirm it with <enter>.
The next position is now blinking. Enter the next letter here,
confirm it with <enter> etc.
If the identifier is complete, quit the text input with <QUIT>.
Confirm the value with <enter>. The method is stored in the
memory.
The method runs now under the identifier "HCl".
A list of contents of the method memory can be printed with
the key sequence <print><user meth><enter>.
>store method
method name: HCl
DET pH D0a HCl
method name: HClmnopq
method name: HIJKLMNO
method name: ABCDEFGH
method name: ********
'um
736 GP Titrino 02103 736.0010
date 1997-09-13 time 07:49
user methods bytes
DET pH Titer_pH D0 190
DET U Blank D0 178
DET U Chloride D0 234
MET U Diazo D0 208
MET Ipol Br-Index D0 226
DET U Ca-Mg D0 278
DET U EDTA-NTA D0 216
DET U Metals D0 190
DET U Perox.No D0 172
DET pH HCl D0 108
remaining bytes 7330
------------
mode and
measured
quantity
method
name
dosing unit, that
is being used