2.1
Survey of the titration modes
736 GP Titrino: Titration, dosing and measuring modes
13
If you are looking for more information to the different quanti-
ties, you will find the display text as a keyword in the index of
the Instructions for Use.
SET, KFT
Endpoint Titration
STAT
Holding a measured value
DOC
Controlled dosing
Titration to preset end-point.
The reagent addition is con-
trolled such that a preset
measured value is adhered to
as accurately as possible.
The dosing function is con-
trolled such that a nominal
measured value is adhered to
as accurately as possible. The
nominal measured value fol-
lows a linear ramp.
Volume that has been dis-
pensed up to the endpoint
(EPX in ml).
Evaluation of
•
rates dV/dt in the set time
windows and the mean
rate over the entire meas-
uring point list.
•
volumes associated with a
preset time (fix V).
•
times associated with a
preset fraction of the total
volume (fix time).
The total volume (C41) and the
time needed for dosing (C42)
can be used for calculation.
•
For rapid, quantitative
determinations in analyti-
cal chemistry. Require-
ment: EP of the titration
reaction is known and
does not change during a
determination series.
•
If an excess of titrant must
be avoided.
•
KFT
: Determinations of
water contents.
For
•
the controlled dosing to a
preset endpoint in the syn-
thesis laboratory.
•
activity determinations, e.g.
of enzymes.
•
kinetic investigations (first
order kinetics).
Sweeping a measured value
gradient for reaction formula-
tions in the syntheses labora-
tory.
V/mL
U/mV
continuously
Control
range
EP
t/s
V/ml
rate
t/s
U/mV
V/ml
U
set