1.4 Working with several dosing units
736 GP Titrino: Getting started
11
Manual dosing
Manual dosing
Manual dosing can be used
•
to prepare an Exchange Unit (or Dosing Unit, resp.)
•
for quick dosing
If you press the key <DOS>, the dosing unit from the method
is working. This dosing unit appears in the display.
If you wish to dose with another dosing unit, use the key
<prep> to select the dosing unit and press <DOS> to dose.
<prep>
<prep>
<DOS>
<FILL>
<START>
<select>
<START>
<QUIT>
Press the key <prep>. In the display appears
"internal D0 prep", i.e. the dosing unit of the Titrino has been
chosen.
Press the key <prep> again. In the display appears
"external D1: prep".
Press the key <DOS> on the Titrino: The external dosing unit
D1 doses. If you press the key <STOP/FILL> on the Titrino, it
will fill again.
With <START>, the function "prep" is started. This prepares
the dosing units. (The parameters for "prep" are under the key
<configuration>, in the group ">prep.dosing elements").
Attention
: With the function "prep", liquid will be expelled!
If the dosing unit D1 is a Dosino, you can empty the dosing
unit completely (e.g. for cleaning). Choose the function
"empty" with the key <select> and start with <START>.
Attention
: With the function "empty", liquid will be expelled!
Quit the selection of dosing units with <QUIT>.
internal D0 prep
external Dl: empty
external D1: prep