5.3 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
942 Extension Module Vario ChS-PP (2.942.0300)
4 mm
■
Insert the new rotor
into the housing
in such a way
that the tubing connections on the rear of the rotor fit into the
corresponding recesses inside the housing and one of the three
holes of the rotor is visible from below in the slot of the housing
.
NOTICE
The rotor's sealing surface is located approx. 4 mm deep inside the
suppressor drive if the rotor is inserted correctly.
If this is not the case, then the rotor must be moved into the cor-
rect position using careful turning. If the rotor cannot be turned or
removed, it can be moved into the correct position from below by
means of a pointed object (e.g. a screwdriver).
5 Cleaning the new connecting piece
■
Clean the sealing surface of the new connecting piece
ethanol using a lint-free cloth.
6 Inserting the new connecting piece
See also Chapter 3.4.1, Page 21
■
into the housing in such a way
that connector
1
is on top and the three pins of the connecting
piece fit into the corresponding recesses on the housing.
■
and tighten by hand (do not use a
tool).
7 Connecting and conditioning the Metrohm Suppressor Mod-
ule (MSM)
■
Reconnect all capillaries of the MSM to the IC system.
■
Before switching the MSM over for the first time, rinse the three
suppressor units with solution for five minutes.