5.2 Maintenance by the user
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
NOTICE
Spare part
A new spare lamp is available from your Metrohm representative under
the article number 6.7430.050.
■
We recommend keeping spare lamps in stock.
■
Only use original lamps in the instrument.
1 Switching off the instrument
■
to the position
O
.
2 Disconnecting the instrument from the energy supply
■
Disconnect the power cord.
■
Wait 10 to 15 minutes until the lamp has cooled down.
3 Opening the lamp compartment
■
Open the lid of the lamp compartment
■
Take the hex key out of the guide rail.
■
Unscrew the four sealing plate screws using the hex key and put
them aside.
■
Remove the sealing plate and put it aside.