3 Installation
■■■■■■■■
17
3 Connecting the flow cell
■
There is an inner thread at the inlet and the outlet of the flow cell.
On each inner thread, mount a tubing with a matching outer
thread.
4 Inserting the flow cell in the DS2500 Holder for flow cells
■
The flow cell has a window. The holder has openings.
Insert the flow cell in the holder so that the window of the flow
cell is positioned at the openings of the holder.
5 Mounting the flow cell spring assembly
■
The holder has 2 openings for the positioning springs of the
spring assembly. Place the spring assembly on the holder. The fol-
lowing is to be observed:
– The 2 tubings of the flow cell are in the center of the
bridge.
– The 2 positioning springs are in the openings of the holder.
■
Fasten the screws on the positioning springs.
6 Connecting the flow cell
■
Connect the inlet of the flow cell to a sample changer.
■
Connect the outlet of the flow cell to the waste container.