3.2 Carrying out titrations
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
870 KF Titrino plus
■
Select
on
and confirm with
[OK]
.
6 Save the method
■
The modified parameters are saved automatically. It is not neces-
sary to save the method under a name of its own.
7 Switch to the main dialog
■
Repeated pressing on
[BACK]
will switch you to the main dialog.
3.2.2
Preparing the buret unit
The
PREP
function is used to rinse the cylinder and tubings of the buret
unit and fill them bubble-free. You should carry out this function before
the first determination or once per day.
Proceed as follows:
1 Open the manual control
■
In the main dialog, select
Menu
and confirm with
[OK]
.
The main menu opens.
■
Select
Manual control
and confirm with
[OK]
.
2 Carry out the PREP dosing function
Note
Make sure that the buret tip is directed into a vessel that can
accommodate the cylinder volume of your buret unit several times
over.