2.2 Setting up the KF titration cell
■■■■■■■■■■■■■■■■■■■■■■
8
■■■■■■■■
870 KF Titrino plus
Equipping the KF titration cell
6.2730.020
6.2730.010
6.0338.100
6.1403.040
6.2104.020
1
2
3
4
5
6
The KF titration cell can now be equipped as follows:
1
Introduce the septum stopper 6.2730.020 (with septum inserted)
into the front opening of the vessel lid.
2
Insert the O-rings of the electrode and of the drying tube into the
middle openings of the vessel lid.
3
Screw the two screw nipples 6.2730.010 into the openings with the
O-rings. Do not screw them in too tightly.
4
Introduce the double Pt-wire electrode 6.0338.100 into the left-hand
opening and then tighten the screw nipple until it seals.
5
Screw the electrode cable 6.2104.020 tightly onto the electrode.
6
Introduce the filled drying tube 6.1403.040 on the right of the elec-
trode into the remaining opening and then tighten the screw nipple
until it seals.