![Metrohm 2.870.0010 Installation And Tutorial Download Page 16](http://html1.mh-extra.com/html/metrohm/2-870-0010/2-870-0010_installation-and-tutorial_1779535016.webp)
2.3 Setting up the aspiration and solvent bottle
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
870 KF Titrino plus
Metrohm offers suitable thread adapters for bottles with other threads
than GL 45.
6.1608.023
6.1819.030
6.1602.105
6.1446.040
6.1805.200
6.1609.010
6.2023.020
6.1801.120
1
2
3
4
5
6
7
To equip the solvent bottle, proceed step-by-step as follows:
1
Place a thread stopper 6.1446.040 in the M6 connector (smallest
opening) of a bottle top 6.1602.105 and screw it tightly.
2
Insert the long PTFE cannula 6.1819.030 into the M8 connector (sec-
ond-smallest opening) of the bottle top.
3
Insert a PTFE tubing 6.1805.200 into the M8 connector of the bottle
top and screw it tightly.
4
Connect a PVC tubing 6.1801.120 to the adsorber tube 6.1609.010
that is filled with the molecular sieve.
5
Place the adsorber tube with the SGJ 14 standard ground joint on
the bottle top.
6
Secure the SGJ 14 standard ground joint of the adsorber tube with
the SGJ clip 6.2023.020.