12
12. . GLP Functions
GLP Functions
726 Titroprocessor, Short Introduction
45
45
12.5.1 Defining a sensor list
Press the <GLP> key on the main page and then the [Sensors] softkey.
ÚÄÄÄ GLP ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ ³
³ Method Validations Date Interval Result ³
³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³
³ 1. ÚÄÄÄ Sensors ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
³ 2. ³ ³ ³
³ 3. ³ Sensor ID Manuf.no. Part.no. Date Quality ³ ³
³ 4. ³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³ ³
³ 5. ³ 1.comb. pH I 0023 0842 6.233.100 96-11-11 OK ³ ³
³ ³ 2.comb. pH II 0023 0941 6.233.100 ³ ³
³ ³ 3. ³ ³
³ ³ 4. ³ ³
³ ³ 5. ³ ³
³ ³ 6. ³ ³
³ ³ 7. ³ ³
³ ³ 8. ³ ³
³ ³ 9. ³ ³
³ ³10. ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³Limits ³ ³ ³ ³Extend.³ ³Delete ³ ³ ³ ³Sensor ³
³ ³ ³ ³ ³info ³ ³sensor ³ ³ ³ ³ test ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
Ten sensors can be defined in a list. The test data of each listed sensor will
be stored and may be viewed and updated at any time.
To define the limits press [Limits]. In the appearing dialog window you can
adjust the limits to the applications this specific sensor is used for. If a sen-
sor is not used for measurements but for titrations only, the limits may be
set less restrictively. The screen depicted below shows the recommended
default values for common applications. These settings may be recalled at
any time by pressing the [Set default] softkey.
ÚÄÄÄ GLP ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ ³
³ Method Validations Date Interval Result ³
³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³
³ 1. ÚÄÄÄ Sensors ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
³ 2. ³ ³ ³
³ 3. ³ Sensor ID Manuf.no. Part.no. Date Quality ³ ³
³ 4. ³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³ ³
³ 5. ³ 1.comb. pH I 0023 0842 6.233.100 96-11-11 OK ³ ³
³ ³ 2.comb. pH II 0023 0941 6.233.100 ³ ³
³ ³ 3. ÚÄÄÄ Sensor Limits ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³ ³
³ ³ 4. ³ ³ ³ ³
³ ³ 5. ³ User def. limits of: comb. pH II ³ ³ ³
³ ³ 6. ³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³ ³ ³
³ ³ 7. ³ slope min. 0.95 ³ ³ ³
³ ³ 8. ³ Zero point U(as) max. 30.0 mV ³ ³ ³
³ ³ 9. ³ Flow potential max. 2.5 mV ³ ³ ³
³ ³10. ³ Response time max. 30 s ³ ³ ³
³ ³ ³ ³ ³ ³
³ ³ ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ ³ ³
³ ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³ ³ ³ ³ ³ Set ³ ³ ³ ³ ³ ³Cancel ³
³ ³ ³ ³ ³default³ ³ ³ ³ ³ ³ ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ