Chapter II
30 MHz oscilloscope
23
Power supply
Mains:
universal mains 105 to 240 Vrms, 48 Hz to 400 Hz
Consumption:
35 W maximum
Removable mains power supply cable
Electromagnetic compatibility
Transmission according to EN 61326-1 (1997) + A1 (1998)
Immunity according to EN 61326-1 (1997) + A1 (1998)
Safety
According to IEC 61010-1 1993, + A2 (1995) (NFC 42020 ; VDE 0411)
Overvoltage categories:
CAT
II
inputs,
300 V
CAT
II
power supply, 264 V
Environment
Usage
indoors
Elevation
< 2 000 m
Insulation
class 1
Degree of pollution
2
Reference temperature
+ 18°C
to + 28°C
Usage temperature
+ 10°C
to + 40°C
Operating temperature
0°C
to + 40°C
Storage temperature
- 20°C
to + 70°C
Relative humidity
< 80% RH
up to 31°C
Mechanical properties
Stackable apparatus with handle used as stand
Mass:
≈
5.5 kg
Packaging
Dimensions:
550 x 460 x 280 mm Mass:
≈
7 kg
7.
SUPPLIES AND OPTIONS
Delivered with the apparatus :
Operating instructions
Mains power supply cable european mains (depending on model)
1/10, 250 MHz, 600 V attenuating probe (2 pieces) (depending on model)
Delivered on request: :
1/1-1/10, 25-150 MHz, 600 V attenuating probe
HX0003
1/10, 250 MHz, 600 V attenuating probe
HX0004
1/100, 200 MHz, 2 kV attenuating probe
HA1317
1/10, 450 MHz, 600 V attenuating probe
HA1323
30 MHz, 700 V differential probe
MX9003
100 A, 100 kHz AC/DC current clamp for oscilloscope
AM0030N
600 A, 10 kHz AC/DC current clamp for oscilloscope
AM0031N
Isolated cable, 50
Ω
BNC / 2 male banana, 1 m (10 pieces)
AG1042
Isolated cable, 50
Ω
BNC / BNC, 1 m (10 pieces)
AG1044
Safety adapter male BNC / 2 female banana (2 pieces)
HA2053
1 male / 2 female BNC-T (3 pieces)
HA2004
50
Ω
BNC passage load
PA4119-50
Carrying bag
AE0189
Dimensions: