Figure 18: MX2034 Scale Factor Validation Set Up
4.1.1.3 ELECTRICALLY ZERO THE PROBE
Set the probe gap “electrically” to the start of the measurement range by observing the DC
output voltage and adjusting the probe position until the output is -1 V +/- 0.1 V at 10 mils
gap.
Figure 19: Setting Electrical Zero
Set micrometer for 10 mils
Move probe until meter reads -1.0V