Technical characteristics of the MX 5006
24
Benchtop multimeters, 6000 and 60000 points
Technical characteristics of the MX 5006 (continued)
V
AC
RMS
Range
Operating
range
Specified
measurement
range
3)
Resolution
Uncertainty
(
±
)
Additional
uncertainty
F (Hz)
1)
Bandwidth
@ 1kHz
Input
impedance
// < 50 pF
Peak
factor
600mV
0 to
600.0mV
60.0 to
600.0mV
0.1mV
1% L +
0.25% x
[F(kHz)-1]L
±5 D
10Hz to
50kHz
10.9M
Ω
3 to
500mV
6V
0 to
6.000V
0.600 to
6.000V
0.001V 10.9M
Ω
3 to 5V
60V
0 to
60.00V
6.00 to
60.00V
0.01V 10.082M
Ω
3 to 50V
600V
0 to
600.0V
60.0 to
600.0V
0.1V 10.008M
Ω
3 to
500V
1000V
2)
0 to
1000V
60 to
1000V
1V
1% L +
0.18% x
[F(kHz)-
1]L ±3 D
45<F<65Hz
0.3% L
typ.
at 100Hz
0.7% L
typ.
at 150Hz
1.8% L
typ.
at 300Hz
30% L
typ.
10Hz to
100kHz
10.008M
Ω
1.42 to
1000V
1) See the typical curve of the 300Hz filter on p. 29.
2) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
3) From 1kHz, the measurement must exceed 15% of the range
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, PEAK
AC and DC voltage
AC+DC TRMS
Range
Operatin
g range
Specified
measurement
range
3)
Reso-
lution
Uncertainty
DC (
±
)
Uncertainty
AC (
±
)
Additional
uncertainty
F (Hz)
1)
Band-
width
Input
impedance
// < 50 pF
Peak
factor
600mV
0 to
600.0mV
60.0 to
600.0mV
0.1mV
0.8% L +
0.18% x
[F(kHz)-1]
L ±5 D
10Hz
to
50kHz
10.9M
Ω
3 to
500mV
6V
0 to
6.000V
0.600 to
6.000V
0.001V
10.9M
Ω
3 to
5V
60V
0 to
60.00V
6.00 to
60.00V
0.01V 10.082M
Ω
3 to
50V
600V
0 to
600.0V
60.0 to
600.0V
0.1V 10.008M
Ω
3 to
500V
1000V
2)
0 to
1000V
60 to 1000V
1V
0.8% L
±10 D
0.8% L +
0.18% x
[F(kHz)-1]
L
±3 D
45<F<65Hz
0.3% L
typ.
at 100Hz
0.7% L
typ.
at 150Hz
1.8% L typ.
at 300Hz
30% L typ.
10Hz
to
100kHz
10.008M
Ω
1.42 to
1000V
1) See the typical curve of the 300Hz filter on p. 29.
2) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
3) From 1kHz, the measurement must exceed 15% of the range
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, PEAK