Technical characteristics of the MX 5006
Benchtop multimeters, 6000 and 60000 points
23
Technical characteristics of the MX 5006
Accuracy:
"n% + nD" means
"n% of the reading
+ n Digit"
(cf. CEI485)
Only values with tolerances or limits are guaranteed values.
Values without tolerances are given for guidance (standard NFC42670).
The technical specifications are guaranteed only after 30 minutes of warming up. Except
as otherwise indicated, they are valid from 5% to 100%
of the measurement range.
VOLTAGES
Protection: 1414Vpk
DC voltage
V
DC
Range
Specified measurement
range
Resolution
Intrinsic error
600mV
0 to 600.0mV
0.1mV
0.5% L + 2 D
6V
0 to 6.000V
0.001V
60V
0 to 60.00V
0.01V
600V
0 to 600.0V
0.1V
1000V
∗
0 to 1000V
1V
0.09% L + 2 D
(
∗
) The display indicates "+OL" above +1050V and "-OL" below -1050V.
AC voltage
V
LowZ AC
RMS
The bandwidth is reduced to 300Hz if the filter is activated. The frequency measurement
is made like a measurement in a 300Hz PB.
1) See the typical curve of the 300Hz filter on p. 29.
2) The LCD indicates "+OL" above +1050V, "-OL" below -1050V or above 1050VRMS.
3) From 1kHz, the measurement must exceed 15% of the range
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, PEAK
Range
Operating
range
Specified
measurement
range
3)
Resolution
Uncertainty
(
±
)
Additional
uncertainty
F (Hz)
1)
Input
impedance
//< 50 pF
Peak
factor
600mV
0 to
600.0mV
60.0 to
600.0mV
0.1mV
1.2% L+
0.25% x
[F(kHz)-1] L
±5 D
3 to
500mV
6V
0 to
6.000V
0.600 to
6.000V
0.001V
3 to 5V
60V
0 to
60.00V
6.00 to
60.00V
0.01V
3 to 50V
600V
0 to
600.0V
60.0 to
600.0V
0.1V
3 to 500V
1000V
2)
0 to
1000V
60 to 1000V
1V
1.2% L+
0.18% x
[F(kHz)-1] L
±3 D
45<F<65Hz
0.3% L
typ.
at 100Hz
0.7% L
typ.
at 150Hz
1.8% L
typ.
at 300Hz
30% L
typ.
≅
520k
Ω
1,42 to
1000V