II - 20
IX 3131
4.2.
Electrical specifications
Accuracy :
@ 23° C ± 5° C ; < 75 % H. R.
Resistance
Range
Max Display
Accuracy test
freq. = 120 Hz
Accuracy test
freq. = 1 kHz
Specified Note
10 Mohm
9.999 M
not specified
+/-(0.6%+5dgts)
after open cal.
1 Mohm
999.9 k
+/-(0.3%+3dgts)
+/-(0.3%+3dgts)
after open cal.
100 kohm
99.99 k
+/-(0.3%+2dgts)
+/-(0.3%+2dgts)
-
10 kohm
9.999 k
+/-(0.3%+2dgts)
+/-(0.3%+2dgts)
-
1 kohm
999.9 ohm
+/-(0.3%+2dgts)
+/-(0.3%+2dgts)
-
100 ohm
99.99 ohm
+/-(0.5%+3dgts)
+/-(0.5%+3dgts)
after short cal.
10 ohm
9.999 ohm
+/-(0.6%+5dgts)
+/-(0.6%+5dgts)
after short cal.
"
"
"
"
Notes :
This specification is based on the measurement performed at the
test socket.
DUT & Test lead to be properly shielded to GUARD if necessary.
Capacitance
Test frequency 120 Hz
Cap. Range
Max Display
Cx Accuracy
DF Accuracy
Specified
Note
10 mF
9.99 mF
(*)
+/-(2.5%+5dgt)
(DF<0.1)
+/-
(5%+100/Cx+5dgt)
(DF<0.1)
after short cal.
1000 µF
999.9 µF
+/-(0.6%+5dgt)
(DF<0.1)
+/-
(1%+100/Cx+5dgt)
(DF<0.1)
after short cal.
100 µF
99.99 µF
+/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
10 µF
9.999 µF
+/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
1000 nF
999.9 nF
+/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
100 nF
99.99 nF
+/-(0.4%+5dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
after open cal.
10 nF
9.999 nF
+/-(0.6%+5dgt)
(DF<0.1)
+/-
(1%+100/Cx+5dgt)
(DF<0.1)
after open cal.
(*) This reading can be extended to 1999 max display.
Summary of Contents for IX 3131
Page 2: ......
Page 4: ...I 2 IX 3131...
Page 6: ...I 4 IX 3131...
Page 24: ...Copyright IX 3131 BENCH TYPE L C R METER Operating manual Chapter II...
Page 25: ...II 2 IX 3131...
Page 27: ...II 4 IX 3131...
Page 31: ...II 8 IX 3131...
Page 42: ...IX 3131 II 19 1 user s manual...
Page 47: ...Copyright IX 3131 RLC MESSBR CKE Bedienungsanleitung Kapitel III...
Page 48: ...III 2 IX 3131...
Page 50: ...III 4 IX 3131...