TeraOhm 5kV
8
3. Warnings
In order to reach the highest level of operator’s safety while carrying out various
measurements and tests using the
TeraOhm 5kV
Tester, as well as to ensure that the
test equipment remains undamaged, it is necessary to consider the following warnings:
MEANING OF SYMBOLS
♦
!
symbol on the instrument means “Read the User Manual with special care!”
♦
symbol on the instrument means “Hazardous voltage higher than 1000 V
may be present at the test terminals!”!
GENERAL PRECAUTIONS
♦
If the test equipment is used in a manner not specified in this Users Manual,
the protection provided by the equipment may be impaired!
♦
Do not use the instrument and accessories, if any damage is noticed!
♦
Consider all generally known precautions in order to avoid the risk of electric
shock while dealing with electric installations!
♦
Service intervention or recalibration procedure can be carried out only by a
competent and authorized person!
♦
Only adequately trained and competent persons may operate the instrument.
BATTERIES
♦
Disconnect all test leads and switch
the
power off before opening
the
Battery
cover!
♦
Avoid connection to main supply when alkaline batteries are inserted, danger
of explosion! Do not charge when alkaline batteries are fitted!
EXTERNAL VOLTAGES
♦
Do not connect the instrument to a mains voltage different from the one
defined on the label adjacent to
the
mains connector, otherwise the instrument
may be damaged.
•
Do not connect test terminals to an external voltage higher than 600 V AC or
DC (CAT III environment) to prevent any damage on the test instrument!
WORKING WITH THE INSTRUMENT
♦
Use only standard or optional test accessories supplied by your distributor!
♦
Equipment under test must be switched off i.e. de-energized before test leads
are connected to the equipment.
♦
Do not touch any conductive parts of equipment under test during the test.
♦
Make sure that the tested object is disconnected (mains voltage disconnected)
before starting the Insulation Resistance measurement!
•
Do not touch
the tested object whilst testing it, risk of electric shock!