MI 3300 PAT Demoboard
Safety warnings
5
2. Safety warnings
•
Only qualified personnel who are familiar with the demoboard and measuring
instruments may use PAT Demoboard!
•
Use only original demonstration tables provided by the manufacturer!
•
The PAT Demoboard must be considered as a real piece of electrical
equipment. The simulated faults are real ones. Despite of the fact that the
simulated leakage currents are kept at relatively low levels it is strictly
forbidden to touch any of the accessible metal parts inside the simulation area
during the demonstration!
•
The PAT Demoboard will work only if connected to a properly earthed TN or
TT outlet. The Ready lamp must lit green for normal operation of the
Demoboard.
If red light is blinking and buzzer sounds immediately disconnect the
demoboard from the mains and check the supply connections !
•
Special care must be taken if performing HV (withstanding) tests. All safety
measures must be considered as if testing real electrical equipment.
•
Always check that the current of HV test equipment is limited to a low level
(<3.5mA) when performing withstanding tests on PAT Demoboard.
•
If the equipment is not used in a manner specified by manufacturer, the
protection provided by equipment may be impaired.
•
Use of Demonstration board in a way not specified in this User Manual could
damage the board.
•
Do not use Demonstration board in case of any damage noticed!
•
Only an authorised person may carry out servicing of Demonstration board!