MI 3300 PAT Demoboard
Demonstration tables information
12
5.5. Portable appliance of class 1 (extension drum)
Simulates
Error value
Error off Error on
Simulation between contacts
Measurement taken
Error off Error on
Error 1
Loose contact of PE conductor
Error value
0,02 Ω
0,84 Ω
PE connection on output socket (1) and
PE connection on mains socket (2)
Earth Bond
0,02 Ω
0,84 Ω
Error 2
Insulation fault
Error value
>20 MΩ
106 kΩ
L and PE pins on socket (1)
Insulation 500 V DC
>20 M
Ω
0,106 M
Ω
Leakage 230 V, 50 Hz
0,00 mA
2,17 mA
Subleakage
0,00 mA
2,17 mA
Error 3
Insulation fault
Error value
30 MΩ
238 kΩ
Isolated accessible metal part (3) and L,
pin on socket (2)
Insulation 500V DC
(probe)
13 MΩ
0,237 MΩ
Touch Leakage 230 V AC 0,00 mA
0,97 mA
Subleakage (probe)
0,00 mA
0,97 mA
Table 5.4: Description of simulated faults / relevant tests