MD 9250 Clamp Meter Series 

Safety 

 

1 Safety 

 

MaintenanceThis  manual  contains  information  and  warnings  that  must  be  followed  for 
operating  the  instrument  safely  and  maintaining  the  instrument  in  a  safe  operating 
condition.  If  the instrument  is used in a manner not  specified by the manufacturer, the 
protection provided by the instrument may be impaired.  
 
The meter protection rating, against the users, is double insulation per IEC/EN61010-1 
2nd Ed., UL61010-1 2nd Ed., CAN/CSA C22.2 No. 61010.1-0.92, IEC/EN61010-2-032, 
UL61010B-2-032, 

CAN/CSA 

C22.2 

No. 

61010-2-032-04 

IEC/EN61010-

031:2002/A1:2008:  
 
Measurement Category IV 1000V AC & DC. 

 
PER IEC61010 OVERVOLTAGE INSTALLATION CATEGORY 
 
OVERVOLTAGE CATEGORY II 

Equipment  of 

OVERVOLTAGE  CATEGORY  II

  is  energy-consuming  equipment  to  be 

supplied from the fixed installation. 
Note 

– Examples include household, office, and laboratory appliances. 

 

OVERVOLTAGE CATEGORY III 

Equipment of 

OVERVOLTAGE CATEGORY III

 is equipment in fixed installations. 

Note 

–  Examples  include  switches  in  the  fixed  installation  and  some  equipment  for 

industrial use with permanent connection to the fixed installation. 

 
OVERVOLTAGE CATEGORY IV
 

Equipment of 

OVERVOLTAGE CATEGORY IV

 is for use at the origin of the installation. 

Note 

–  Examples  include  electricity  meters  and  primary  over-current  protection 

equipment. 
 

TERMS IN THIS MANUAL 
WARNING

  identifies  conditions  and  actions  that  could  result  in  serious  injury  or  even 

death to the user. 
 

CAUTION 

identifies conditions and actions that could cause damage or malfunction in 

the instrument. 
 

WARNING

  

To reduce the risk of fire or electric shock, do not expose this product to rain or moisture. 
The meter is intended only for indoor use. 
 
To avoid  electrical shock hazard,  observe the proper safety precautions when working 
with  voltages  above  60  VDC  or  30  VAC  rms.  These  voltage  levels  pose  a  potential 
shock hazard to the user.  
 
Keep your hands/fingers behind the hand/finger barriers (of the meter and the test leads) 
that  indicate  the  limits  of  safe  access  of  the  hand-held  part  during  measurement. 
Inspect  test  leads,  connectors,  and  probes  for  damaged  insulation  or  exposed  metal 
before using the instrument. If any defects are found, replace them immediately. 

Summary of Contents for MD 9250

Page 1: ...METREL MD 9250 Industrial TRMS AC DC CAT IV 1000 V MD 9250 User Manual Bedienungsanleitung Version 1 1 Code no 20 752 003...

Page 2: ...Test Measurement Unit 16 1st Qtr Business Park Blenheim Road Epsom Surrey KT19 9QN Great Britain E mail info metrel co uk Internet http www metrel co uk 2012 2016 METREL Mark on your equipment certifi...

Page 3: ...ntents Inhalt English 1 Safety 4 2 Cenelec Directives 6 3 Product Description 7 4 Operation 8 5 Specifications 16 6 Maintenance 21 Deutsch 1 Sicherheitsbestimmungen 23 2 Cenelec Richtlinien 24 3 Produ...

Page 4: ...es in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation OVERVOLTAGE CATEGORY IV Equipment of OVERVOLTAGE CATEGORY IV is for use at the or...

Page 5: ...where measurement is to be carried out could be accessible CAUTION Disconnect the test leads from the test points before changing meter functions INTERNATIONAL ELECTRICAL SYMBOLS Caution Refer to the...

Page 6: ...MD 9250 Clamp Meter Series Cenelec Directives 6 2 Cenelec Directives The instruments conform to CENELEC Low voltage directive 2006 95 EC and Electromagnetic compatibility directive 2004 108 EC...

Page 7: ...ry switch Selector to turn the power ON OFF and Select a function 4 Push buttons for special functions features 5 Input Jack for all functions EXCEPT non invasive DCA ACA current functions 6 Common Gr...

Page 8: ...the resistance is below the Audible Threshold the meter further gives a continuity beep tone When a signal above the voltage threshold of 1 5V DC or AC up to the rated 1000V is present the meter displ...

Page 9: ...be up to 566mA 1000V x 1 414 2 5k decreasing abruptly to approx 3 37mA 1000V x 1 414 420k within a fraction of a second Do not use AutoCheckTM mode on circuits that could be damaged by such low input...

Page 10: ...than one pole of a circuit may result in differential current like identifying leakage current measurement Locate the conductor s at the Jaws center as much as possible to get the best measuring accu...

Page 11: ...sed across the diode The digital display shows OL if the diode is good Any other readings indicate the diode is resistive or shorted defective CAUTION 1 Using Resistance Continuity Diode or Capacitanc...

Page 12: ...AMD 9023 with correct polarities You can also use a plug adapter AMD 9024 Optional purchase with banana pins to type K socket to adapt other type K standard mini plug temperature probes Electric Field...

Page 13: ...cation Optional purchase PC interface kit AMD 9250 is required to connect the meter to the PC computer RS232 or USB ports Press and hold the HOLD button while turning the meter on to enable meter PC C...

Page 14: ...hold the button for 1 second or more and release to resume auto ranging Note Manual ranging feature is not available in Hz and function ranges Set Beeper Off Press the RANGE button while turning the...

Page 15: ...ing mode where available The annunciator turns off Display Backlight Press the SELECT button for 1 second or more to toggle the display backlight on and off Auto Power Off APO When the meter is on the...

Page 16: ...ions 1000V rms E M C Meets EN61326 1 2006 EN55022 EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000 4 8 EN61000 4 11 In an RF field of 3V m Capacitance functi...

Page 17: ...on sinusoidal waveforms DC Voltage RANGE Accuracy 6 000V 60 00V 600 0V 1000V 0 5 5d Input Impedance 10M 50 pF nominal AutoCheckTM_ DCV RANGE Accuracy 6 000V 60 00V 600 0V 1000V 1 3 5d AutoCheckTM Lo Z...

Page 18: ...bove 50V typical Ended up impedances vs display voltages typically are 10k 100V 60k 300V 200k 600V 420k 1000V VFD_ACV with Low Pass Filter RANGE Accuracy 1 10Hz 20Hz 6 000V 60 00V 600 0V 1000V 4 80d 2...

Page 19: ...000V 1 0 3d Test Current 0 56mA typically Open Circuit Voltage 1 8VDC typically DCA Current Clamp on RANGE Accuracy 1 2 200 0A 2 0 5d 0 500A 2 0 5d 500 2000A 3 0 5d 1 Induced error from adjacent curr...

Page 20: ...readings if any Temperature RANGE Accuracy 50 oC 1000 oC 0 3 4d 58 oF 1832 oF 0 3 6d K type thermocouple range accuracy not included Hz Line Level Frequency Function Sensitivity Sine RMS Range 6V 2V...

Page 21: ...usible resistors will be blown off become high impedance like fuses to protect the user and the instrument Most measuring functions through this terminal will then be open circuit The series fusible r...

Page 22: ...specifications of the equipment or the user manual Damage after a drop a shock or flooding The contents of this manual must not be reproduced in any form whatsoever without the consent of the manufac...

Page 23: ...tionen Hinweis Beispiele sind Schalter in festen Installationen und einige Anlagen im Industriegebrauch mit permanentem Anschluss an die feste Installation BERSPANNUNGSKATEGORIE IV Ger te nach BERSPAN...

Page 24: ...etroffen werden wenn Benutzer bei der Messung mit stromf hrenden Teilen in Ber hrung kommen k nnten ACHTUNG Trennen Sie die Pr fkabel von den Pr fpunkten bevor Sie die Messfunktionen ndern INTERNATION...

Page 25: ...Drehschalter zum Ein Ausschalten und Ausw hlen der Funktion 4 Drucktasten f r Sonderfunktionen 5 Eingangsbuchse f r alle Funktionen AUSSER nicht invasiver AC und DC Strom messfunktionen 6 Gemeinsame...

Page 26: ...er H rschwelle liegt gibt das Messger t einen durchgehenden Piepton aus Wenn ein Signal ber der Spannungsschwelle von 1 5 V DC oder AC bis zum Nennwert von 1000 V anliegt zeigt das Messger t den entsp...

Page 27: ...nnerhalb von Sekundenbruchteilen auf 3 37 mA 1000 V x 1 414 420 k abfallen Verwenden Sie nicht den AutoCheckTM Modus an Stromkreisen da diese durch eine derart geringe Eingangsimpedanz besch digt werd...

Page 28: ...DC A und DC AC A Hz Die Eingabe erfolgt f r nicht invasive Strommessungen ber die Stromzangenbacken Standardeinstellung auf ACA Hz Funktion Dr cken Sie die Taste SELECT und l sen Sie sie wieder um di...

Page 29: ...en Kontinuierliche Messung Kapazit ts und Diodenfunktion Die Eingabe erfolgt ber die Pr fleitungsanschl sse Standardeinstellung auf Kapazit t Dr cken Sie die Taste SELECT und lassen Sie sie wieder los...

Page 30: ...MD 9250 Stromzange Serie Betrieb 30 Temperaturfunktion...

Page 31: ...gelangen Das Messger t zeigt E F an sobald es bereit ist Die Signalst rke wird in mehreren Balkendiagramm Segmenten auf dem Display und zus tzlichen Piept nen angezeigt Kontaktfreie EF Erfassung Am ob...

Page 32: ...Taste SELECT mindestens eine Sekunde gedr ckt um die Hintergrundbeleuchtung des Displays ein oder auszuschalten Die Hintergrundbeleuchtung wird nach 32 Sekunden wieder ausgeschaltet um die Batteriela...

Page 33: ...rte als Referenzwerte f r aufeinanderfolgende Messungen zu bestimmen Das Display zeigt die Messwerte nun im Vergleich zu den gespeicherten Referenzwerten an Das hei t Anzeige Messwert gespeicherter We...

Page 34: ...2 50 s Spitzen berlastschutz Zange angebracht 2000 A Eff fortlaufend und COM Anschl sse alle anderen Funktionen 1000 V Eff EMV Erf llt EN61326 1 2006 EN55022 EN61000 3 2 EN61000 3 3 EN61000 4 2 EN6100...

Page 35: ...t H chstwert Crest Faktor 1 4 1 bei voller Skala und 2 8 1 bei halber Skala sowie mit Frequenzkomponenten innerhalb der spezifizierten Frequenzbandbreite f r Nicht Sinus Wellenformen Gleichspannung Me...

Page 36: ...normal bersteigt M gliche Impedanzwerte sind bei jeweiligen Spannungsgr en 10 k 100 V 60 k 300 V 200 k 600 V 420 k 1000 V VFD_ACV mit Tiefpassfilter Messbereich Genauigkeit 1 10 Hz 20 Hz 6 V 60 V 600...

Page 37: ...normal Leerlaufspannung 1 8 V DC normal Gleichstrom Zange angelegt Messbereich Genauigkeit 1 2 200 A 2 0 5 d 0 500 A 2 0 5 d 500 2000 A 3 0 5 d 1 Induzierter Fehler von n chstem stromf hrenden Leiter...

Page 38: ...tur Messbereich Genauigkeit 50 oC 1000 oC 0 3 4 d 58 oF 1832 oF 0 3 6 d Bereich und Genauigkeit f r Thermoelement Typ K nicht vorhanden Hz Bezugspegelfrequenz Funktion Empfindlichkeit Sinus Effektivwe...

Page 39: ...hnlich einer Sicherung ausgel st Hochimpedanz um den Benutzer und das Messger t zu sch tzen Die meisten Messfunktionen dieses Anschlusses laufen dann im offenen Schaltkreis Die Schmelzwiderst nde und...

Page 40: ...ie ausgeschlossen sind Unsachgem er Gebrauch wie z B berlastung des Ger tes oder Verwendung von nicht zugelassenen Zubeh r Gewaltanwendung Besch digung durch Fremdeinwirkungen oder durch Fremdk rper z...

Reviews: