A 1199 

ρ

-Adapter 

ρ

-Adapter operation 

 

10 

 

4. 

ρ

-Adapter operation 

 

ρ

-Adapter 1

st

 use and later battery replacement 

 

Required tool: screwdriver 

 

ρ

-Adapter must be disconnected from any device. 

Put the 

ρ

-Adapter on desktop with backside up (

figure 3.3

). 

Remove screws (

pos. 4

 on 

figure 3.3

). 

Remove battery cover. 

Insert battery cells correctly in battery holder (

figure 3.4

). 

Return back the battery cover and screw it. 

Press a key on the 

ρ

-adapter. LED (

pos. 1

 on 

figure 3.2

) must light. 

The 

ρ

-Adapter is ready for application. 

 

ρ

-Adapter application 

 

Prepare test setup for the measurement of specific earth resistance. 

Connect test leads of the 

ρ

-Adapter to the spikes, inserted into the ground plane. 

Connect the 

ρ

-Adapter to the instrument and turn its power on. 

On the instrument select 

EARTH

 test function and 

ρ

 subfunction and enter the 

distance 

a

 between spikes. 

Start measurement(s) and store result(s) (optional) after finished measurement. 

If necessary rearrange spikes and repeat the measurements. 

After finished measurements, turn power off of the 

ρ

-Adapter and the instrument. 

Disconnect test leads, remove inserted spikes, clean them and keep all in 
accessory bag. 

 

Figure 4.1: Connection circuit for measurement of specific earth resistance 

Specific earth resistance is calculated according to the following formula: 
 

ρ

 = 2

π

aR 

with: 

ρ 

Specific earth resistance [

Ω

m (

Ω

ft)]. 

The distance between spikes [m (ft)]. 

Measured earth resistance. 

 

Summary of Contents for A1199

Page 1: ...METREL test and measurement accessories Adapter A1199 Instruction manual Version 1 1 Code no 20 751 041...

Page 2: ...l metrel si Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations 2006 METREL No part of...

Page 3: ...afety and operational considerations 5 1 1 Warnings and notes 5 1 2 Battery 5 1 3 Standards applied 6 3 Adapter description 7 3 1 Adapter set 9 4 Adapter operation 10 5 Maintenance 11 5 1 Cleaning 11...

Page 4: ...d using METREL Adapter accessory The Adapter is special accessory intended for using only with METREL installation testers and designed for simple measurement with attached instrument In general the A...

Page 5: ...ce intervention The Adapter contains alkaline or NiMH rechargeable battery It should only be replaced with the same type as defined on the battery placement label or in this manual All normal safety p...

Page 6: ...ral requirements Electromagnetic compatibility EMC EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements Class B Hand held equipment used in controlled EM environme...

Page 7: ...onnection to measuring connector of test equipment 1 2 3 Figure 3 1 Adapter parts Legend 1 Connector plate for 4 wire universal cable connection Do not connect measuring wires to any external voltage...

Page 8: ...r 1 LED Indicates battery condition Continuous lighting battery good Blinking weak battery 2 Power Switches the Adapter power on and off 1 4 2 3 Figure 3 3 Adapter back site Legend 1 Back side label 2...

Page 9: ...AA alkaline or rechargeable NiMH battery 3 1 Adapter set Adapter Universal test cable 4 wire Extension cable red 15 m Two earthing spikes Small soft carrying bag Note Earth 3 wire 20 m set S2026 is al...

Page 10: ...test leads of the Adapter to the spikes inserted into the ground plane Connect the Adapter to the instrument and turn its power on On the instrument select EARTH test function and subfunction and ente...

Page 11: ...ument 5 2 Service For repairs under warranty or at any other time please contact your distributor 6 Technical specifications 6 1 General data Power supply voltage 6 VDC 4 x 1 5 V alkaline or 4 x 1 2 V...

Reviews: