A 1199 

ρ

-Adapter 

Safety and operational considerations 

 

2. Safety and operational considerations 

 

1.1  Warnings and notes 

 
In order to reach high level of operator’s safety while carrying out measurements with 
the 

ρ

-Adapter A1199 as well as to keep the test accessory and equipment undamaged, 

it is necessary to consider the following general warnings:  

 

 

 Warning on the 

ρ

-Adapter means »Read the Instruction manual with 

special care to safety operation«. The symbol requires an action!  

 

If the 

ρ

-Adapter is used in a manner not specified in this user manual or the 

manual of target test equipment, the protection provided by the 

ρ

-Adapter 

and equipment may be impaired! 

 

Read this user manual carefully, otherwise use of the 

ρ

-Adapter may be 

dangerous for the operator, for test equipment or for the tested object! 

 

Do not use the 

ρ

-Adapter if any damage is noticed! 

 

Never use the 

ρ

-Adapter in energized installation! 

 

Only a competent authorized person should carry out service intervention! 

 

The 

ρ

-Adapter contains alkaline or NiMH rechargeable battery. It should 

only be replaced with the same type as defined on the battery placement 
label or in this manual. 

 

 

All normal safety precautions have to be taken in order to avoid risk of 
electric shock when working on electrical installations!

 

 

1.2 Battery  

 
The 

ρ

-Adapter uses 4 pieces of size AA 1.5 V alkaline or 1.2 V NiMH battery cells. 

Battery condition is always present by its LED on front side of the adapter when it is 
turned on. 
Blinking LED means weak battery. 

 

Notes

 

 Disconnect the 

ρ

-Adapter form tested object and form the instrument, and 

turn its power OFF before opening battery cover. 

 

All four cells should be replaced each time all of the same type and capacity! 

 

If 

ρ

-Adapter is not used for longer period, remove its battery. 

 

Insert battery cells correctly, otherwise the 

ρ

-Adapter will not operate and the 

battery could be discharged.  

 

Take into account handling, maintenance and recycling requirements that are 
defined by related regulations and manufacturer of alkaline or rechargeable 
batteries! 

 

Turn the 

ρ

-Adapter power off after finished measurement to save the battery. 

Summary of Contents for A1199

Page 1: ...METREL test and measurement accessories Adapter A1199 Instruction manual Version 1 1 Code no 20 751 041...

Page 2: ...l metrel si Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations 2006 METREL No part of...

Page 3: ...afety and operational considerations 5 1 1 Warnings and notes 5 1 2 Battery 5 1 3 Standards applied 6 3 Adapter description 7 3 1 Adapter set 9 4 Adapter operation 10 5 Maintenance 11 5 1 Cleaning 11...

Page 4: ...d using METREL Adapter accessory The Adapter is special accessory intended for using only with METREL installation testers and designed for simple measurement with attached instrument In general the A...

Page 5: ...ce intervention The Adapter contains alkaline or NiMH rechargeable battery It should only be replaced with the same type as defined on the battery placement label or in this manual All normal safety p...

Page 6: ...ral requirements Electromagnetic compatibility EMC EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements Class B Hand held equipment used in controlled EM environme...

Page 7: ...onnection to measuring connector of test equipment 1 2 3 Figure 3 1 Adapter parts Legend 1 Connector plate for 4 wire universal cable connection Do not connect measuring wires to any external voltage...

Page 8: ...r 1 LED Indicates battery condition Continuous lighting battery good Blinking weak battery 2 Power Switches the Adapter power on and off 1 4 2 3 Figure 3 3 Adapter back site Legend 1 Back side label 2...

Page 9: ...AA alkaline or rechargeable NiMH battery 3 1 Adapter set Adapter Universal test cable 4 wire Extension cable red 15 m Two earthing spikes Small soft carrying bag Note Earth 3 wire 20 m set S2026 is al...

Page 10: ...test leads of the Adapter to the spikes inserted into the ground plane Connect the Adapter to the instrument and turn its power on On the instrument select EARTH test function and subfunction and ente...

Page 11: ...ument 5 2 Service For repairs under warranty or at any other time please contact your distributor 6 Technical specifications 6 1 General data Power supply voltage 6 VDC 4 x 1 5 V alkaline or 4 x 1 2 V...

Reviews: